CLASS I Device recall · Reported 8 May 2013 · FDA Enforcement Report

GemStar Pump - Battery recalled by Hospira

Hospira is recalling GemStar Pump - Battery, distributed nationwide in the US. The recall was initiated on 18 Mar 2013 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesList Numbers: 13000, 13100, 13150
Quantity recalledList Number 13000: 25,119 pumps; List Number 13100: 185 pumps; List Number 13150: 13,497 pumps

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1159-2013
FDA event ID
64839
Recalling firm
Hospira, Lake Forest, IL
Classification
Class I
Status
Terminated
Recall initiated
18 Mar 2013
FDA classified
26 Apr 2013
Reported
8 May 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

18 Mar 2013Recall initiated by company
26 Apr 2013Classified by FDA+39 days
8 May 2013Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Battery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

GemStar Pump - Battery Product Usage: The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients.

Where it was distributed

Worldwide Distribution - USA (nationwide) Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.

Nationwide

Questions about this recall

Is GemStar Pump - Battery recalled?

Yes. Hospira recalled GemStar Pump - Battery on 18 Mar 2013. The recall is Class I and its status is Terminated. Recall number Z-1159-2013.

Why was it recalled?

Battery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.

Which lots are affected?

List Numbers: 13000, 13100, 13150

Where was it sold?

Worldwide Distribution - USA (nationwide) Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Hospira

All 225

Other Other Medical Devices recalls

All 13,778