CLASS II Device recall · Reported 31 Aug 2016 · FDA Enforcement Report

The HomeChoice automated peritoneal dialysis system recalled

Baxter Healthcare is recalling The HomeChoice automated peritoneal dialysis system, distributed nationwide in the US. The recall was initiated on 7 Oct 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes*** 1) HomeChoice automated peritoneal dialysis system: Product Codes: 5C4471 and 5C4471R; Serial Numbers: All serial numbers; *** 2) HomeChoice Pro automated peritoneal dialysis system: Product Codes: 5C8310 and 5C8310R; Serial Numbers: All serial numbers
Quantity recalled*** 1) Product Codes 5C4471 and 5C4471R: Approximately 48,600 units; *** 2) Product Codes 5C8310 and 5C8310R: Approxiamtely 16,990 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2609-2016
FDA event ID
66606
Recalling firm
Baxter Healthcare, Deerfield, IL
Classification
Class II
Status
Terminated
Recall initiated
7 Oct 2013
FDA classified
22 Aug 2016
Reported
31 Aug 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Oct 2013Recall initiated by company
22 Aug 2016Classified by FDA+1050 days
31 Aug 2016Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Where it was distributed

Worldwide distribution. US nationwide including Puerto Rico; Australia, Canada, New Zealand, and Japan

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites.

Which lots are affected?

*** 1) HomeChoice automated peritoneal dialysis system: Product Codes: 5C4471 and 5C4471R; Serial Numbers: All serial numbers; *** 2) HomeChoice Pro automated peritoneal dialysis system: Product Codes: 5C8310 and 5C8310R; Serial Numbers: All serial numbers

Where was it sold?

Worldwide distribution. US nationwide including Puerto Rico; Australia, Canada, New Zealand, and Japan

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Baxter Healthcare

All 573