CLASS II ONGOING Device recall · Reported 14 Aug 2024 · FDA Enforcement Report

Hookups provide the interface between an Advantage Plus" recalled

Steris is recalling Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor, distributed nationwide in the US. The recall was initiated on 13 Jun 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: UDI-DI 00677964091223; Lot Numbers 404360, 415307, 421967, 423540, 425801, 427239, 435309, 436068, 447102, 516866, 523989, 539379, 540599, 542619, 544723, 550675, 554204, 560426, 587537, 588583, 594221, 604965, 61450
Quantity recalled37

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2536-2024
FDA event ID
94987
Recalling firm
Steris, Mentor, OH
Classification
Class II
Status
Ongoing
Recall initiated
13 Jun 2024
FDA classified
8 Aug 2024
Reported
14 Aug 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

13 Jun 2024Recall initiated by company
8 Aug 2024Classified by FDA+56 days
14 Aug 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection

Product as listed by the FDA

Hookup Model/Catalog Number: 2-8-540 Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes.

Where it was distributed

Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Malaysia, Qatar, Ukraine, United Arab Emirates, Nicaragua, and Nigeria.

Nationwide

Stores named: Online Retailers

Questions about this recall

Is Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor recalled?

Yes. Steris recalled Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor on 13 Jun 2024. The recall is Class II and its status is Ongoing. Recall number Z-2536-2024.

Why was it recalled?

IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection

Which lots are affected?

Lot Code: UDI-DI 00677964091223; Lot Numbers 404360, 415307, 421967, 423540, 425801, 427239, 435309, 436068, 447102, 516866, 523989, 539379, 540599, 542619, 544723, 550675, 554204, 560426, 587537, 588583, 594221, 604965, 61450

Where was it sold?

Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Malaysia, Qatar, Ukraine, United Arab Emirates, Nicaragua, and Nigeria.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Steris

All 113

Other Surgical Instruments recalls