CLASS II Device recall · Reported 14 Mar 2018 · FDA Enforcement Report
i-STAT DE handheld data processing module for clinical use recalled
Abbott Point Of Care is recalling i-STAT DE handheld data processing module for clinical use, distributed in 49 states. The recall was initiated on 12 Sep 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0926-2018
- FDA event ID
- 79150
- Recalling firm
- Abbott Point Of Care, Princeton, NJ
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 12 Sep 2017
- FDA classified
- 8 Mar 2018
- Reported
- 14 Mar 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 12 Sep 2017 | Recall initiated by company | |
| 8 Mar 2018 | Classified by FDA | +177 days |
| 14 Mar 2018 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)
Where it was distributed
Distributed in 48 States: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, Australia, Italy, Sweden, and Canada.
AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland
Questions about this recall
Why was it recalled?
Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.
Which lots are affected?
All units that have been upgraded to software version 2.8
Where was it sold?
Distributed in 48 States: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, Australia, Italy, Sweden, and Canada.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Abbott Point Of Care
All 31| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | i-STAT EG6+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
| Device | CLASS II | i-STAT EG7+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
| Device | CLASS II | i-STAT G3+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
| Device | CLASS II | i-STAT EG7+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
| Device | CLASS II | i-STAT CG8+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
Other Medical Software recalls
All 930| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | SmartPath to dStream for XR and 3.0T recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Upgrade to MR recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Positron Emission Tomography and Computed Tomography System recalled | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Tandem t:slim mobile application t:slim Mobile Application, iOS recalled | Tandem Diabetes Care | Device Malfunction | Nationwide |
| Device | CLASS II | Ingenia Elition S and Ingenia Elition recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |