CLASS II ONGOING Device recall · Reported 21 Jun 2023 · FDA Enforcement Report

Icelock 125 Ratchet recalled by Ossur H / F

Ossur H / F is recalling Icelock 125 Ratchet, distributed in 44 states. The recall was initiated on 12 May 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF/UDI-DI: L-125000/05690967566142; L-621200/05690967596590; L-621000/05690967118334; L-621100/05690967155322; L-692020/05690967128548. Lots: HF210205, HF210519, HF211008, HF220303, MX210518, MX210816, MX210823, MX211022, MX220316.
Quantity recalled4,190

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1906-2023
FDA event ID
92207
Recalling firm
Ossur H / F, Reykjavik, N/A
Classification
Class II
Status
Ongoing
Recall initiated
12 May 2023
FDA classified
13 Jun 2023
Reported
21 Jun 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Lead & Heavy Metals

Timeline

12 May 2023Recall initiated by company
13 Jun 2023Classified by FDA+32 days
21 Jun 2023Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of the external prosthetic system that replaces a missing lower limb, which can lead to falling and injury.

Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.

Product as listed by the FDA

Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock 621 Ratchet Adaption Kit, REF: L-621100; Ratchet Lock Body, REF: L-692020

Where it was distributed

US: SC, FL,NM, PA, VA, CA, TX, MT, UT, OH, NV, KS, MI, GA, IA, WA, TN, AZ, MN, NC, IN, MS, AL, NY, ID, DE, OK, MA, WI, OR, KY, MD, CO, WV, LA, IL, NJ, CT, MO, NE, WY, AR, SD, PR

AlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesota

Questions about this recall

Is Icelock 125 Ratchet recalled?

Yes. Ossur H / F recalled Icelock 125 Ratchet on 12 May 2023. The recall is Class II and its status is Ongoing. Recall number Z-1906-2023.

Why was it recalled?

A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of the external prosthetic system that replaces a missing lower limb, which can lead to falling and injury.

Which lots are affected?

REF/UDI-DI: L-125000/05690967566142; L-621200/05690967596590; L-621000/05690967118334; L-621100/05690967155322; L-692020/05690967128548. Lots: HF210205, HF210519, HF211008, HF220303, MX210518, MX210816, MX210823, MX211022, MX220316.

Where was it sold?

US: SC, FL,NM, PA, VA, CA, TX, MT, UT, OH, NV, KS, MI, GA, IA, WA, TN, AZ, MN, NC, IN, MS, AL, NY, ID, DE, OK, MA, WI, OR, KY, MD, CO, WV, LA, IL, NJ, CT, MO, NE, WY, AR, SD, PR

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ossur H / F

All 29

Other Other Medical Devices recalls

All 13,778