CLASS II ONGOING Device recall · Reported 22 Apr 2026 · FDA Enforcement Report

iLet Ace Pump Kit recalled by Beta Bionics

Beta Bionics is recalling iLet Ace Pump Kit, distributed nationwide in the US. The recall was initiated on 13 Aug 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesILet Ace Pumps containing software version BOM000036 Rev C (version 1.4.2) or older.
Quantity recalled15 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1809-2026
FDA event ID
98528
Recalling firm
Beta Bionics, Irvine, CA
Classification
Class II
Status
Ongoing
Recall initiated
13 Aug 2025
FDA classified
10 Apr 2026
Reported
22 Apr 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

13 Aug 2025Recall initiated by company
10 Apr 2026Classified by FDA+240 days
22 Apr 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software

Where it was distributed

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, MA, MD, ME, MI, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY AND WASHINGTON DC.

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaIdahoIllinoisIndianaLouisianaMassachusettsMarylandMaineMichiganMinnesotaMissouri

Questions about this recall

Why was it recalled?

it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.

Which lots are affected?

ILet Ace Pumps containing software version BOM000036 Rev C (version 1.4.2) or older.

Where was it sold?

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, MA, MD, ME, MI, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY AND WASHINGTON DC.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Beta Bionics

All 3
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIiLet Bionic Pancreas recalled by Beta BionicsBeta BionicsDevice Malfunction51 states
DeviceCLASS IIiLet Bionic Pancreas recalled by Beta BionicsBeta BionicsDevice MalfunctionNationwide

Other Medical Software recalls

All 932