CLASS II Device recall · Reported 10 Jun 2020 · FDA Enforcement Report

Implantable Insulin Pump System recalled by Medtronic

Medtronic is recalling Medtronic Implantable Insulin Pump System. The recall was initiated on 10 Apr 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesDevice not distributed within the US. All Pump lots produced May 21, 2019 to present containing the following serial numbers: 31359, 31360, and 31332
Quantity recalled24 Pumps

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2206-2020
FDA event ID
85582
Recalling firm
Medtronic, Northridge, CA
Classification
Class II
Status
Terminated
Recall initiated
10 Apr 2020
FDA classified
2 Jun 2020
Reported
10 Jun 2020
Type
Voluntary: Firm initiated
Customers notified by
Telephone
Reason category
Sterility

Timeline

10 Apr 2020Recall initiated by company
2 Jun 2020Classified by FDA+53 days
10 Jun 2020Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001

Where it was distributed

No US Distribution. OUS: Netherlands, France, and Sweden

Questions about this recall

Is Implantable Insulin Pump System recalled?

Yes. Medtronic recalled Implantable Insulin Pump System on 10 Apr 2020. The recall is Class II and its status is Terminated. Recall number Z-2206-2020.

Why was it recalled?

Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death.

Which lots are affected?

Device not distributed within the US. All Pump lots produced May 21, 2019 to present containing the following serial numbers: 31359, 31360, and 31332

Where was it sold?

No US Distribution. OUS: Netherlands, France, and Sweden

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.