CLASS II Device recall · Reported 27 Apr 2016 · FDA Enforcement Report

IntelliSpace Portal DX/HX/EX AutoSPECT Pro recalled

Philips Medical Systems (Cleveland) is recalling IntelliSpace Portal DX/HX/EX AutoSPECT Pro, distributed in 16 states. The recall was initiated on 4 Mar 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesM/N 881001; S/N: 85254, 82202, 85330, 82117, 7000022, 105009, 970001, 850001, 870030, 82141, 95068, 87182, 87197, 95078, 82140, 11017, 98139, 95035, 87200, 980047, 85314, 85303, 980046, 85298, 822022, 7000019, 950076, 950115, 95043, 95047, 950021, 960123 & 105006.
Quantity recalled33 Units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1506-2016
FDA event ID
73776
Recalling firm
Philips Medical Systems (Cleveland), Cleveland, OH
Classification
Class II
Status
Terminated
Recall initiated
4 Mar 2016
FDA classified
15 Apr 2016
Reported
27 Apr 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

4 Mar 2016Recall initiated by company
15 Apr 2016Classified by FDA+42 days
27 Apr 2016Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.

Where it was distributed

Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.

ColoradoDelawareFloridaGeorgiaIllinoisIndianaKansasMissouriNorth CarolinaNevadaNew YorkOhioOregonPennsylvaniaTennesseeTexas

Questions about this recall

Why was it recalled?

The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.

Which lots are affected?

M/N 881001; S/N: 85254, 82202, 85330, 82117, 7000022, 105009, 970001, 850001, 870030, 82141, 95068, 87182, 87197, 95078, 82140, 11017, 98139, 95035, 87200, 980047, 85314, 85303, 980046, 85298, 822022, 7000019, 950076, 950115, 95043, 95047, 950021, 960123 & 105006.

Where was it sold?

Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.