CLASS II ONGOING Device recall · Reported 19 Dec 2018 · FDA Enforcement Report
IntelliVue MX40 Patient Monitor recalled by Philips Electronics North
Philips Electronics North America is recalling IntelliVue MX40 Patient Monitor, distributed nationwide in the US. The recall was initiated on 10 Oct 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0592-2019
- FDA event ID
- 81531
- Recalling firm
- Philips Electronics North America, Andover, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 10 Oct 2018
- FDA classified
- 10 Dec 2018
- Reported
- 19 Dec 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Patient Monitors
Timeline
| 10 Oct 2018 | Recall initiated by company | |
| 10 Dec 2018 | Classified by FDA | +61 days |
| 19 Dec 2018 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
IntelliVue MX40 Patient Monitor, Model Nos. 865350, 865351, 865352, 867146 - Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals.
Where it was distributed
US Nationwide distribution.
Questions about this recall
Why was it recalled?
A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
Which lots are affected?
All devices manufactured or repaired from 21-FEB-2017 to 22-AUG -2018. (Serial numbers included as an attachment; too many to list.)
Where was it sold?
US Nationwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips Electronics North America
All 160| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips HeartStart FR3 Defibrillator recalled | Philips Electronics North America | Device Malfunction | Nationwide |
| Device | CLASS II | Allura Xper F010, System recalled by Philips Electronics North | Philips Electronics North America | Other | Nationwide |
| Device | CLASS II | Allura Xper FD2O Biplane, System recalled by Philips Electronics North | Philips Electronics North America | Other | Nationwide |
| Device | CLASS II | AIIuraXperFDlO, System recalled by Philips Electronics North | Philips Electronics North America | Other | Nationwide |
| Device | CLASS II | Integris Allura 9 F Fdxd, System recalled by Philips Electronics North | Philips Electronics North America | Other | Nationwide |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |