CLASS II ONGOING Device recall · Reported 5 Oct 2022 · FDA Enforcement Report

Leica Surgical Operating Microscopes, Arveo recalled

Leica Microsystems is recalling Leica Surgical Operating Microscopes, Arveo, distributed nationwide in the US. The recall was initiated on 26 Aug 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: 210721001, 180721001, 190721001, 121021001, 100821001, 60821001, 90821001, 120821001, 10821001, 80721001, 40821001, 130821001, 290821001, 240821001, 91021001, 180921001, 181121001, 151021001, 190821001, 160921001, 20721001, 50721001, 40921001, 150821001, 160821001
Quantity recalled25 devices

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1849-2022
FDA event ID
90818
Recalling firm
Leica Microsystems, deerfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
26 Aug 2022
FDA classified
29 Sep 2022
Reported
5 Oct 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

26 Aug 2022Recall initiated by company
29 Sep 2022Classified by FDA+34 days
5 Oct 2022Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.

Product as listed by the FDA

Leica Surgical Operating Microscopes, Arveo, Part Number 10448999

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.

Which lots are affected?

Serial Numbers: 210721001, 180721001, 190721001, 121021001, 100821001, 60821001, 90821001, 120821001, 10821001, 80721001, 40821001, 130821001, 290821001, 240821001, 91021001, 180921001, 181121001, 151021001, 190821001, 160921001, 20721001, 50721001, 40921001, 150821001, 160821001

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

3 other products · FDA event 90818

This product is part of a recall event; the main page for the event is Leica Surgical Operating Microscopes, M530 OHX recalled.

More recalls from Leica Microsystems

All 40

Other Surgical Instruments recalls

All 3,883