CLASS II Device recall · Reported 21 Dec 2016 · FDA Enforcement Report

Lumex, Patriot Full Electric Homecare Bed recalled by GF Health Products

GF Health Products is recalling Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed, distributed nationwide in the US. The recall was initiated on 30 Sep 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesBED SERIAL NUMBER RANGE US0458 / US0468 Patriot Full Electric 120V 697142145108 through 697142147927 US0458 / US0468 Patriot Full Electric 220V 697144140132 through 697144140172 US6000 Patriot LX Full Electric 120V 69204141262 through 69204141737 SERVICE MOTOR KITS WITH SHIP DATES OF MAY 4, 2015 THROUGH JUNE 30, 2016 Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX Full Electric 120V 690-8004-912
Quantity recalled3,547 (3063 beds, 484 motors)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0732-2017
FDA event ID
75460
Recalling firm
GF Health Products, Fond Du Lac, WI
Classification
Class II
Status
Terminated
Recall initiated
30 Sep 2016
FDA classified
12 Dec 2016
Reported
21 Dec 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

30 Sep 2016Recall initiated by company
12 Dec 2016Classified by FDA+73 days
21 Dec 2016Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000. The failure mode is directly related to contaminated material in the plastic housing used to produce the DUOMAT 3 motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design. If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed containing the Okin Dumat 3 Head/Foot Motors. Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX Full Electric 120V 690-8004-912

Where it was distributed

Worldwide Distribution - USA ( nationwide) Distribution including Puerto Rico and to the countries of : Bahamas, Canada, Chile, El Salvador and Santo Domingo.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000. The failure mode is directly related to contaminated material in the plastic housing used to produce the DUOMAT 3 motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design. If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported.

Which lots are affected?

BED SERIAL NUMBER RANGE US0458 / US0468 Patriot Full Electric 120V 697142145108 through 697142147927 US0458 / US0468 Patriot Full Electric 220V 697144140132 through 697144140172 US6000 Patriot LX Full Electric 120V 69204141262 through 69204141737 SERVICE MOTOR KITS WITH SHIP DATES OF MAY 4, 2015 THROUGH JUNE 30, 2016 Description Part Number Motor and Pendant Patriot Full Electric 120V 690-3001-943 / 690-7001-943 Boxed Full Electric 220V Motor 690-3220-943 Motor and Pendant Patriot LX Full Electric 120V 690-8004-912

Where was it sold?

Worldwide Distribution - USA ( nationwide) Distribution including Puerto Rico and to the countries of : Bahamas, Canada, Chile, El Salvador and Santo Domingo.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from GF Health Products

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIDrop Arm Versamode" recalled by Gf Health ProductsGf Health ProductsOtherNationwide

Other Other Medical Devices recalls

All 13,778