CLASS II ONGOING Device recall · Reported 10 Jun 2026 · FDA Enforcement Report

MAC 5 A4 Resting ECG Analysis System recalled

GE Medical Systems China is recalling MAC 5 A4 Resting ECG Analysis System 855001-001, distributed nationwide in the US. The recall was initiated on 21 Apr 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware version 3.00 855001-001, UDI-DI 00195278276070 8855002-001, UDI-DI 00195278276124
Quantity recalled16

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2306-2026
FDA event ID
98917
Recalling firm
GE Medical Systems China, Wuxi
Classification
Class II
Status
Ongoing
Recall initiated
21 Apr 2026
FDA classified
3 Jun 2026
Reported
10 Jun 2026
Type
Voluntary: Firm initiated
Reason category
Device Malfunction

Timeline

21 Apr 2026Recall initiated by company
3 Jun 2026Classified by FDA+43 days
10 Jun 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph

Where it was distributed

Worldwide - US Nationwide distribution in the states of Florida and Georgia. The countries of India, Italy, Republic of Korea, and United Kingdom.

Nationwide FloridaGeorgia

Questions about this recall

Why was it recalled?

Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.

Which lots are affected?

Software version 3.00 855001-001, UDI-DI 00195278276070 8855002-001, UDI-DI 00195278276124

Where was it sold?

Worldwide - US Nationwide distribution in the states of Florida and Georgia. The countries of India, Italy, Republic of Korea, and United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from GE Medical Systems China

All 25

Other Surgical Instruments recalls

All 3,891