CLASS I ONGOING Device recall · Reported 15 Apr 2026 · FDA Enforcement Report

Medline medical convenience kits recalled by Medline Industries

Medline Industries is recalling Medline medical convenience kits, containing Namic Angiographic Rotating Adaptor (RA), distributed nationwide in the US. The recall was initiated on 27 Feb 2026 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesMedline SKU # DYNJ63021B, UDI/DI each 10193489833775UDI/DI case 40193489833776, Lot Number: 26AMD887; Medline SKU # DYNJ64297B, UDI/DI each 10198459472992UDI/DI case 40198459472993, Lot Number: 26AMA970; Medline SKU # DYNJ82411, UDI/DI each 10195327193232UDI/DI case 40195327193233, Lot Number: 25IMB049.
Quantity recalled20 kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1722-2026
FDA event ID
98598
Recalling firm
Medline Industries, Northfield, IL
Classification
Class I
Status
Ongoing
Recall initiated
27 Feb 2026
FDA classified
7 Apr 2026
Reported
15 Apr 2026
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Other Reasons

Timeline

27 Feb 2026Recall initiated by company
7 Apr 2026Classified by FDA+39 days
15 Apr 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Product as listed by the FDA

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. LAPAROSCOPIC HERNIORRHAPHY DSC, Medline SKU # DYNJ63021B; 2. LAP PACK, Medline SKU # DYNJ64297B; 3. MINOR PROCEDURE PACK, Medline SKU # DYNJ82411.

Where it was distributed

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Nationwide CaliforniaPuerto Rico

Questions about this recall

Is Medline medical convenience kits, containing Namic Angiographic Rotating Adaptor (RA) recalled?

Yes. Medline Industries recalled Medline medical convenience kits, containing Namic Angiographic Rotating Adaptor (RA) on 27 Feb 2026. The recall is Class I and its status is Ongoing. Recall number Z-1722-2026.

Why was it recalled?

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Which lots are affected?

Medline SKU # DYNJ63021B, UDI/DI each 10193489833775UDI/DI case 40193489833776, Lot Number: 26AMD887; Medline SKU # DYNJ64297B, UDI/DI each 10198459472992UDI/DI case 40198459472993, Lot Number: 26AMA970; Medline SKU # DYNJ82411, UDI/DI each 10195327193232UDI/DI case 40195327193233, Lot Number: 25IMB049.

Where was it sold?

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Medline Industries

All 686

Other Surgical Instruments recalls

All 3,809