CLASS I Device recall · Reported 8 May 2013 · FDA Enforcement Report

Medtronic DBS Therapy for Dystonia Kit recalled

Medtronic Neuromodulation is recalling Medtronic DBS Therapy for Dystonia Kit, distributed nationwide in the US. The recall was initiated on 8 Feb 2013 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesall product is affected as the lead kits are not being removed from the market. Medtronic is providing Additional Instructions for Capping the Lead and Removing the Lead Cap.
Quantity recalled30,000

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1191-2013
FDA event ID
64439
Recalling firm
Medtronic Neuromodulation, Minneapolis, MN
Classification
Class I
Status
Terminated
Recall initiated
8 Feb 2013
FDA classified
1 May 2013
Reported
8 May 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Lead & Heavy Metals

Timeline

8 Feb 2013Recall initiated by company
1 May 2013Classified by FDA+82 days
8 May 2013Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential for lead damage due to the use of the lead cap provided in DBS and Dystonia kits. Medtronic has received reports of DBS leads being damaged at the connector end of the lead when the lead cap is used. The connector end of the lead is the end of the lead connected to the lead extension. Tightening or loosening of the setscrew may twist the setscrew connector block and may damag

Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.

Product as listed by the FDA

Medtronic DBS Therapy for Dystonia Kit, models 3317, 3319, 3337 and 3339. Sterile and Non-Pyrogenic. Product Usage: Dystonia Therapy Kit is indicated for unilateral or bilateral stimulation of the internal globus pallidus (GPi) or the subthalamic nucleus (STN) to aid in the management of chronic, intractable (drug refractory) primary dystonia, including generalized and/or segmental dystonia, hemidystonia, and cervical dystonia (torticollis).

Where it was distributed

Worldwide Distribution - USA Nationwide and the countries of: Algeria, Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Viet Nam.

Nationwide Puerto Rico

Questions about this recall

Is Medtronic DBS Therapy for Dystonia Kit recalled?

Yes. Medtronic Neuromodulation recalled Medtronic DBS Therapy for Dystonia Kit on 8 Feb 2013. The recall is Class I and its status is Terminated. Recall number Z-1191-2013.

Why was it recalled?

There is a potential for lead damage due to the use of the lead cap provided in DBS and Dystonia kits. Medtronic has received reports of DBS leads being damaged at the connector end of the lead when the lead cap is used. The connector end of the lead is the end of the lead connected to the lead extension. Tightening or loosening of the setscrew may twist the setscrew connector block and may damag

Which lots are affected?

all product is affected as the lead kits are not being removed from the market. Medtronic is providing Additional Instructions for Capping the Lead and Removing the Lead Cap.

Where was it sold?

Worldwide Distribution - USA Nationwide and the countries of: Algeria, Australia, Austria, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Viet Nam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Medtronic Neuromodulation

All 86

Other Other Medical Devices recalls

All 13,778