CLASS II ONGOING Device recall · Reported 27 Oct 2021 · FDA Enforcement Report
Medtronic sterile single-use blades that are components recalled
Medtronic Xomed is recalling Medtronic sterile single-use blades that are components of the StraightShot M Series, distributed nationwide in the US. The recall was initiated on 31 Aug 2021 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0157-2022
- FDA event ID
- 88586
- Recalling firm
- Medtronic Xomed, Jacksonville, FL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 31 Aug 2021
- FDA classified
- 21 Oct 2021
- Reported
- 27 Oct 2021
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 31 Aug 2021 | Recall initiated by company | |
| 21 Oct 2021 | Classified by FDA | +51 days |
| 27 Oct 2021 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
There was an increase in complaints related to tip breaks and wobble/vibration.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Medtronic sterile single-use blades that are components of the StraightShot M Series Handpiece and Fusion ENT Navigation Software, and the Midas Rex Microsaw and Triton System with the following product numbers: (1) REF 1883080EM, Rotatable Fusion Blade, Quadcut 3.0 mm x 13 cm; (2) REF 1883480EM, Rotatable Fusion Blade, Quadcut, 3.4 mm x 13 cm; (3) REF 1884006EM, Rotatable Fusion Blade, RAD 40; (4) REF 1884012EM, Rotatable Fusion Blade, RAD 12; (5) REF 1884080EM, Rotatable Fusion Blade, Tricut, 13 cm; (6) REF 1884380EM, Rotatable Fusion Blade, Quadcut, 4.3 mm x 13 cm.
Where it was distributed
Worldwide distribution - US Nationwide, including PR. There was also government/military distribution and the countries of Australia, Austria, Bahrain, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Republic of Korea, Kuwait, Latvia, Libya, Luxembourg, Macao, Malaysia, Mauritius, Mongolia, Namibia, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Viet Nam.
Questions about this recall
Why was it recalled?
There was an increase in complaints related to tip breaks and wobble/vibration.
Which lots are affected?
(1) REF 1883080EM, GTIN/UPN 00763000034009, Lot/serial numbers: 0217668426, 0217668430, 0217668431, 0218547904, 0219386765, 0219387546, 0221279639, 0221597049; (2) REF 1883480EM, GTIN/UPN 00763000034016, Lot/serial numbers: 0218355352, 0218355353, 0218355354, 0218548106, 0218975242, 0218975243, 0219488206, 0219810719, 0221111103, 0221275012, 0221367934, 0221597050, 0221791924; (3) REF 1884006EM, GTIN/UPN 00763000043254, Lot/serial numbers: 0217260953, 0217260954, 0217260955, 0217260956, 0217260957, 0217260958, 0217260959, 0217969622, 0217969623, 0217969624, 0217969625, 0217970026, 0217970027, 0218162408, 0218162409, 0218282135, 0218282136, 0218282139, 0218311142, 0218311143, 0218311144, 0218311145, 0218356286, 0218356287, 0218356288, 0218356289, 0218356290, 0218356291, 0218356292, 0218975837, 0219215914, 0219215915, 0219215916, 0219715476, 0219715477, 0219811260, 0220659888, 0220804521, 0220826103, 0220878167, 0220878168, 0220929588, 0220996074, 0221199850, 0221367938, 0221477065, 0221477146, 0221586895, 0221618118, 0221791912, 0222046728, 0222046729; (4) REF 1884012EM, GTIN/UPN 00763000043247, Lot/serial numbers: 0217667211, 0217667212, 0217667213, 0217667214, 0217667215, 0217667216, 0218159075, 0218159076, 0218159077, 0218159452, 0218159453, 0218159455, 0218159456, 0218159457, 0218159458, 0218423331, 0218975832, 0219036454, 0219036455, 0219122536, 0219122537, 0221791915; (5) REF 1884080EM, GTIN/UPN 00763000033842, Lot/serial numbers: 0217258057, 0217258058, 0217258059, 0217258060, 0217258061, 0217258062, 0217258063, 0217922715, 0217922716, 0217968723, 0217968724, 0217968725, 0217969086, 0217969087, 0217969088, 0217969089, 0217969090, 0217969091, 0217969092, 0217969093, 0217969094, 0217969095, 0217969096, 0217969098, 0217969099, 0217969100, 0218351460, 0218351462, 0218351463, 0218351464, 0218351465, 0218354786, 0218354787, 0218354788, 0218354789, 0218496725, 0218569925, 0218572166, 0218572167, 0218577805, 0218578866, 0218578867, 0218578868, 0218761931, 0218761932, 0218761933, 0218761934, 0218761935, 0218761936, 0218772873, 0218795848, 0218795849, 0218795850, 0218795851, 0218962903, 0218972892, 0218972893, 0218972894, 0218972895, 0218972896, 0219090839, 0219180699, 0219214780, 0219214781, 0219214782, 0219214783, 0219214784, 0219434676, 0219488207, 0219745076, 0220804698, 0220929480, 0220929481, 0220929482, 0220929483, 0221091281, 0221111100, 0221279635, 0221430167, 0221476225, 0221477046, 0221477047, 0221477048, 0221477049, 0221477051, 0221619326, 0221619330, 0221619331, 0221619332, 0221619333, 0221619334, 0221784366, 0221786297, 0221791918, 0221805157, 0221805158, 0221824122, 0221824123; (6) REF 1884380EM, GTIN/UPN 00763000034023, Lot/serial numbers: 0217258064, 0217258065, 0217260946, 0217260947, 0217260948, 0217260949, 0217260950, 0217260951, 0217260952, 0217262726, 0218179453, 0218348413, 0218355677, 0218355678, 0218355679, 0218355680, 0218355681, 0218355682, 0218355683, 0218355684, 0218355685, 0218422557, 0218422558, 0218496125, 0218497688, 0218497689, 0218497690, 0218572168, 0218683395, 0218683396, 0218683397, 0218683398, 0218795858, 0218795859, 0218795860, 0218975828, 0218975829, 0218975830, 0219215906, 0219215907, 0219215908, 0219215910, 0219653288, 0219653289, 0219802564, 0219834731, 0219848776, 0219854757, 0219890572, 0219890573, 0220484577, 0220484578, 0220659891, 0220659896, 0220703724, 0220703986, 0220703987, 0220703988, 0220703989, 0220878170, 0220878171, 0220878172, 0220878173, 0221072337, 0221110719, 0221110720, 0221118536, 0221118537, 0221118538, 0221118539, 0221181704, 0221182518, 0221271504, 0221278671, 0221279628, 0221279629, 0221305678, 0221319087, 0221468021, 0221477059, 0221477060, 0221489745, 0221489926, 0221611261, 0221616348, 0221619315, 0221619316, 0221619317, 0221619318, 0221792011, 0221805165, 0221805386, 0221805388, 0221824447, 0221824448, 0221824449;
Where was it sold?
Worldwide distribution - US Nationwide, including PR. There was also government/military distribution and the countries of Australia, Austria, Bahrain, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Republic of Korea, Kuwait, Latvia, Libya, Luxembourg, Macao, Malaysia, Mauritius, Mongolia, Namibia, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Viet Nam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
2 other products · FDA event 88586This product is part of a recall event; the main page for the event is Medtronic Sterile single-use blades that are components recalled.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Medtronic Sterile single-use blades that are components recalled | Medtronic Xomed | Device Malfunction | Nationwide | |
| CLASS II | Medtronic sterile single-use blades that are components recalled | Medtronic Xomed | Device Malfunction | Nationwide |
More recalls from Medtronic Xomed
All 25| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | IPC Powerease System recalled by Medtronic Xomed | Medtronic Xomed | Other | 25 states |
| Device | CLASS II | NIM Vital Nerve Monitoring System: Console NIM4CM01 NIM recalled | Medtronic Xomed | Device Malfunction | Nationwide |
| Device | CLASS II | Medtronic ENT Ultra Round Steel Cutting Bur recalled by Medtronic Xomed | Medtronic Xomed | Other | FL |
| Device | CLASS II | NIM TriVantage EMG Endotracheal Tubes recalled by Medtronic Xomed | Medtronic Xomed | Other | FL |
| Device | CLASS I | NIM EMG Endotracheal Tube, Part Numbers: a) recalled by Medtronic Xomed | Medtronic Xomed | Other | Nationwide |
Other Medical Software recalls
All 932| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Kit: Urinary Clean Catch 20/cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Device Malfunction | CA |
| Device | CLASS II | Spectral CT and Spectral CT 7500 RT recalled by Philips Medical Systems | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | SmartPath to dStream for XR and 3.0T recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Upgrade to MR recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Positron Emission Tomography and Computed Tomography System recalled | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |