CLASS II Device recall · Reported 24 Feb 2016 · FDA Enforcement Report
MEE-1000A Neuromaster recalled by Nihon Kohden America
Nihon Kohden America is recalling MEE-1000A Neuromaster, distributed nationwide in the US. The recall was initiated on 19 Jan 2012 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0817-2016
- FDA event ID
- 73114
- Recalling firm
- Nihon Kohden America, Irvine, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 19 Jan 2012
- FDA classified
- 17 Feb 2016
- Reported
- 24 Feb 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 19 Jan 2012 | Recall initiated by company | |
| 17 Feb 2016 | Classified by FDA | +1490 days |
| 24 Feb 2016 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed to prohibit unintentional current flow.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
MEE-1000A Neuromaster, Model No. Accessories: MS-110B (Electric Stimulation), MS-120BK (Current Stimulator Box) JB-116B, JB-132B (Amplifiers) Electromyography. (EMG)
Where it was distributed
Worldwide distribution. US nationwide, Australia, Austria, France, Germany, Greece, Japan, Romania, South Africa, South Korea, Spain, and United Kingdom.
Questions about this recall
Is MEE-1000A Neuromaster recalled?
Yes. Nihon Kohden America recalled MEE-1000A Neuromaster on 19 Jan 2012. The recall is Class II and its status is Terminated. Recall number Z-0817-2016.
Why was it recalled?
The MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed to prohibit unintentional current flow.
Which lots are affected?
Serial Numbers: MS-110B: (17 units) 00013-00019, 00023- 00032 MS-120BK: (6 units) 0009- 00014 JB-116B: (31 units), 0008, 00012, 00014, 00016, 00018, 00019, 00020, 00021, 00023, 00025, 00026, 00027, 00028, 00029, 00031, 00032, 00034, 00035, 00039, 00043, 00102,00103, 00122-00128, 00134, 00135 JB-132B: (83 units) 00009, 00011, 00012, 00014, 00015, 00016, 00018, 00019, 00020, 00021, 00023, 00024, 00025, 00026, 00027, 00036, 00037, 00038, 00041, 00042, 00043, 00044, 00046, 00047, 00049, 00061, 00062, 00066, 00067, 00068, 00069,00071, 00072, 00074, 00075, 00082, 00084, 0085, 00086, 00090, 00095, 00099, 00120, 00121-00129, 00131-00147, 00149-00155, 00157, 00161, 00164-00168
Where was it sold?
Worldwide distribution. US nationwide, Australia, Austria, France, Germany, Greece, Japan, Romania, South Africa, South Korea, Spain, and United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Nihon Kohden America
All 10| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Adult Cap-ONE Biteblock recalled by Nihon Kohden America | Nihon Kohden America | Other | Nationwide |
| Device | CLASS II | Nihon Kohden Adult/Pediatric Forehead Disposable SpO2 Sensor, 9m recalled | Nihon Kohden America | Device Malfunction | 29 states |
| Device | CLASS II | Nihon Kohden Adult Ear Clip SpO2 Sensor, 5 meter recalled | Nihon Kohden America | Device Malfunction | 29 states |
| Device | CLASS II | BSM-3000 Series Bedside Monitor recalled by Nihon Kohden America | Nihon Kohden America | Device Malfunction | Nationwide |
| Device | CLASS II | WMTS Telemetry Receiver, Multiple Patient Receiver recalled | Nihon Kohden America | Other | 46 states |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |