CLASS II Device recall · Reported 15 Jan 2014 · FDA Enforcement Report

Multiples Anspach Micro Curved Attachments (MCA) recalled

The Anspach Effort is recalling Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters, distributed nationwide in the US. The recall was initiated on 26 Oct 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll Lots for MCA - MICRO CURVED ATTACHMENT, MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM, MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM, MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM, MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM, MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM, MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM, MCA5-1SB - lMM FLUTED BALL, 5.6CM, MCA5-1SD - lMM DIAMOND BALL, 5.6CM, MCA5-2SB - 2MM FLUTED BALL, 5.6CM, MCA5-2SD - 2MM DIAMOND BALL, 5.6CM, MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM, MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM, MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM, MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM, MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM, MCA7-15SD  1.5MM DIAMOND BALL, 7.3CM, MCA7-1SB - lMM FLUTED BALL, 7.3CM, MCA7-1SD - 1 MM DIAMOND BALL, 7 .3CM, MCA7-2SB - 2MM FLUTED BALL, 7.3CM and MCA7-2SD - 2MM DIAMOND BALL, 7.3CM.
Quantity recalled6,824 imits (4,7324 in USA / 2,100 Foreign)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0708-2014
FDA event ID
67124
Recalling firm
The Anspach Effort, Palm Beach Gardens, FL
Classification
Class II
Status
Terminated
Recall initiated
26 Oct 2012
FDA classified
9 Jan 2014
Reported
15 Jan 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

26 Oct 2012Recall initiated by company
9 Jan 2014Classified by FDA+440 days
15 Jan 2014Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

During a recent review of documentation Anspach determined that the current directions for use for the MCA cutting burrs and MCA Micro Curved Attachment do not provide clear details regarding the intended use specifically for the MCA burrs. The MCA burrs are intended for use in delicate bone in otologic procedures such as cochleostomy. The use of the device in dense bone removal may result in cu

Product as listed by the FDA

Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters and Attachments) MCA - MICRO CURVED ATTACHMENT, MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM, MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM, MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM, MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM, MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM, MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM, MCA5-1SB - lMM FLUTED BALL, 5.6CM, MCA5-1SD - lMM DIAMOND BALL, 5.6CM, MCA5-2SB - 2MM FLUTED BALL, 5.6CM, MCA5-2SD - 2MM DIAMOND BALL, 5.6CM, MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM, MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM, MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM, MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM, MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM, MCA7-15SD  1.5MM DIAMOND BALL, 7.3CM, MCA7-1SB - lMM FLUTED BALL, 7.3CM, MCA7-1SD - 1 MM DIAMOND BALL, 7 .3CM, MCA7-2SB - 2MM FLUTED BALL, 7.3CM and MCA7-2SD - 2MM DIAMOND BALL, 7.3CM. The MCA attachment is a reusable device used with dissection tools. Dissection tools are single use, disposable cutter designed for cutting and shaping delicate bone, primary in otology procedures for delicate bone cutting.

Where it was distributed

Worldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, Great Britain, France, Sweeden, Italy, Spain, Japan, South Africa, Australia, Switzerland,

Nationwide

Questions about this recall

Is Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters recalled?

Yes. The Anspach Effort recalled Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters on 26 Oct 2012. The recall is Class II and its status is Terminated. Recall number Z-0708-2014.

Why was it recalled?

During a recent review of documentation Anspach determined that the current directions for use for the MCA cutting burrs and MCA Micro Curved Attachment do not provide clear details regarding the intended use specifically for the MCA burrs. The MCA burrs are intended for use in delicate bone in otologic procedures such as cochleostomy. The use of the device in dense bone removal may result in cu

Which lots are affected?

All Lots for MCA - MICRO CURVED ATTACHMENT, MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM, MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM, MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM, MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM, MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM, MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM, MCA5-1SB - lMM FLUTED BALL, 5.6CM, MCA5-1SD - lMM DIAMOND BALL, 5.6CM, MCA5-2SB - 2MM FLUTED BALL, 5.6CM, MCA5-2SD - 2MM DIAMOND BALL, 5.6CM, MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM, MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM, MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM, MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM, MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM, MCA7-15SD  1.5MM DIAMOND BALL, 7.3CM, MCA7-1SB - lMM FLUTED BALL, 7.3CM, MCA7-1SD - 1 MM DIAMOND BALL, 7 .3CM, MCA7-2SB - 2MM FLUTED BALL, 7.3CM and MCA7-2SD - 2MM DIAMOND BALL, 7.3CM.

Where was it sold?

Worldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, Great Britain, France, Sweeden, Italy, Spain, Japan, South Africa, Australia, Switzerland,

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from The Anspach Effort

All 260

Other Other Medical Devices recalls

All 13,771