CLASS II ONGOING Device recall · Reported 14 Oct 2026 · FDA Enforcement Report

Nihon Kohden CNS-6200 Series Central Nurse Station recalled

Nihon Kohden America is recalling Nihon Kohden CNS-6200 Series Central Nurse Station, distributed nationwide in the US. The recall was initiated on 8 Sep 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40
Quantity recalled808

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0089-2027
FDA event ID
99549
Recalling firm
Nihon Kohden America, Irvine, CA
Classification
Class II
Status
Ongoing
Recall initiated
8 Sep 2026
FDA classified
6 Oct 2026
Reported
14 Oct 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

8 Sep 2026Recall initiated by company
6 Oct 2026Classified by FDA+28 days
14 Oct 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No

Where it was distributed

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN. KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NL, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, QA, WA, WI, WV and WY.

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasLouisianaMassachusettsMarylandMaineMichigan

Questions about this recall

Why was it recalled?

Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.

Which lots are affected?

Lot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40

Where was it sold?

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN. KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NL, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, QA, WA, WI, WV and WY.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Nihon Kohden America

All 11

Other Patient Monitors recalls

All 1,022