CLASS II ONGOING Device recall · Reported 18 Jun 2025 · FDA Enforcement Report
OBM00002 OBM DAB (Digital Acquisition Box) recalled by Natus Neurology
Natus Neurology is recalling OBM00002 OBM DAB (Digital Acquisition Box), distributed nationwide in the US. The recall was initiated on 7 May 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1925-2025
- FDA event ID
- 96761
- Recalling firm
- Natus Neurology, Oakville
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 7 May 2025
- FDA classified
- 9 Jun 2025
- Reported
- 18 Jun 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 7 May 2025 | Recall initiated by company | |
| 9 Jun 2025 | Classified by FDA | +33 days |
| 18 Jun 2025 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital Acquisition Box) Software Version: N/A Product Description: OBM00002 OBM DAB (Digital Acquisition Box) Component: No
Where it was distributed
Worldwide distribution - US Nationwide including Puerto Rico (PR) and OUS (foreign) countries of: Italy (IT), Estonia (EE), Israel (IL), China (CN), Germany (DE), Saudi Arabia (SA), Switzerland (CH), Chile (CL), Colombia (CO), Netherlands (the) (NL), Japan (JP), United Kingdom of Great Britain and Northern Ireland(the) (GB), Russian Federation (the)(RU), Bolivia (Plurinational State of) (BO), Canada (CA), France (FR), Peru (PE), Guyana (GY), Mexico (MX), Belgium (BE), Thailand (TH), Paraguay (PY), Hong Kong (HK), United Arab Emirates (AE), Austria (AT), India (IN), Indonesia (ID), Kuwait (KW), Qatar (QA), Lithuania (LT), Denmark (DK), Myanmar (MM), Uruguay (UY), Tunisia (TN), Turkey (TR), Korea (the Republic of) (KR), Hungary (HU), Slovenia (SI), Argentina (AR), Norway (NO), Oman (OM), Poland (PL), Finland (FI), Ireland (IE), Australia (AU), Latvia (LV), Ecuador (EC), Singapore (SG), Spain (ES), Serbia (RS), South Africa (ZA), Romania (RO), Algeria (DZ), Portugal (PT), Dominican Republic (the) (DO), Lebanon (LB), Malaysia (MY), Costa Rica (CR), Viet Nam (VN), & Bahrain (BH).
Nationwide ArkansasCaliforniaColoradoIllinoisPuerto RicoTennessee
Questions about this recall
Why was it recalled?
Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.
Which lots are affected?
UDI-DI 00382830010825 Serial Numbers 001 through OBM00002H3613. Note earlier version serial numbers did not contain the sequence of "OBM"
Where was it sold?
Worldwide distribution - US Nationwide including Puerto Rico (PR) and OUS (foreign) countries of: Italy (IT), Estonia (EE), Israel (IL), China (CN), Germany (DE), Saudi Arabia (SA), Switzerland (CH), Chile (CL), Colombia (CO), Netherlands (the) (NL), Japan (JP), United Kingdom of Great Britain and Northern Ireland(the) (GB), Russian Federation (the)(RU), Bolivia (Plurinational State of) (BO), Canada (CA), France (FR), Peru (PE), Guyana (GY), Mexico (MX), Belgium (BE), Thailand (TH), Paraguay (PY), Hong Kong (HK), United Arab Emirates (AE), Austria (AT), India (IN), Indonesia (ID), Kuwait (KW), Qatar (QA), Lithuania (LT), Denmark (DK), Myanmar (MM), Uruguay (UY), Tunisia (TN), Turkey (TR), Korea (the Republic of) (KR), Hungary (HU), Slovenia (SI), Argentina (AR), Norway (NO), Oman (OM), Poland (PL), Finland (FI), Ireland (IE), Australia (AU), Latvia (LV), Ecuador (EC), Singapore (SG), Spain (ES), Serbia (RS), South Africa (ZA), Romania (RO), Algeria (DZ), Portugal (PT), Dominican Republic (the) (DO), Lebanon (LB), Malaysia (MY), Costa Rica (CR), Viet Nam (VN), & Bahrain (BH).
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Natus Neurology
All 16| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Natus Brain Monitor Breakout recalled by Natus Neurology | Natus Neurology | Device Malfunction | 7 states |
| Device | CLASS II | Sleep Supplies Start Kit with Single Use Supplies recalled | Natus Neurology | Other | Nationwide |
| Device | CLASS II | Tensive Conductive Adhesive Gel recalled by Natus Neurology | Natus Neurology | Other | Nationwide |
| Device | CLASS II | Ipela Fhd Ptz Tile Kits with Axis cameras used recalled | Natus Neurology | Device Malfunction | Nationwide |
| Device | CLASS II | natus newborn care Olympic Brainz Monitor recalled by Natus Neurology | Natus Neurology | Other | Nationwide |
Other Medical Software recalls
All 903| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Brilliance recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips IQon Spectral CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Ingenuity CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Brilliance CT 64: Brilliance CT 64 Channel recalled | Philips North America | Device Malfunction | MA |
| Device | ANNOUNCED | BMC Luna G3 APAP recalled by BMC MedicalBMC | BMC Medical | Device Malfunction | Nationwide |