CLASS II ONGOING Device recall · Reported 27 May 2026 · FDA Enforcement Report

OneLIF Interbody Cage recalled by Novapproach Spine

Novapproach Spine is recalling OneLIF Interbody Cage (, distributed nationwide in the US. The recall was initiated on 23 Apr 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF/UDI-DI/ Cage Lots: 010-107-1007/G07001010710070/ 103003, 105701, 118001, 129301; 010-107-1007-2/G070010107100720/ 114301, 117509, 128703; 010-107-1209/G07001010712090/ 102714, 105702, 117501; 010-107-1411/G07001010714110/ 103019, 105703; 010-107-1613/G07001010716130/ 103020, 105704, 127701; 010-113-1006/G07001011310060/ 103021, 105705, 118002, 129801; 010-113-1006-2/G070010113100620/ 117510, 127716, 114302; 010-113-1208/G07001011312080/ 103022, 105706, 117502, 126902; 010-113-1410/G07001011314100/ 102715, 105707, 127702; 010-113-1612/G07001011316120/ 103023, 105708, 127703; 010-119-1407/G07001011914070/ 103024, 105709, 125101, 128701; 010-119-1609/G07001011916090/ 102716, 105710, 127001; 010-119-1811/G07001011918110/ 103025, 105711, 127003; 010-125-1606/G07001012516060/ 102717, 117901, 127002; 010-125-1808/G07001012518080/ 103005, 117902, 127704; 010-207-1007/G07001020710070/ 102702, 106201, 118003, 125102; 010-207-1007-2/G070010207100720/ 117511, 124106, 127717, 114303; 010-207-1209/G07001020712090/ 102701, 105401, 124101, 127705, 101201; 010-207-1411/G07001020714110/ 105001, 105402, 101203; 010-207-1613/G07001020716130/ 103007, 105403, 124102, 127706; 010-213-1206/G07001021312060/ 105002, 105404, 121202, 124103, 127707, 101204; 010-213-1408/G07001021314080/ 102704, 105405, 121203, 124104, 127708; 010-213-1610/G07001021316100/ 103008, 105406, 130101; 010-213-1812/G07001021318120/ 103009, 105407, 127709; 010-219-1406/G07001021914060/ 102705, 105408, 117503, 119301, 121201, 124105; 010-219-1608/G07001021916080/ 103010, 105409, 121204, 127710; 010-219-1810/G07001021918100/ 103011, 105410, 127711; 010-225-1807/G07001022518070/ 102706, 113701, 117001, 121205; 010-225-2009/G07001022520090/ 103001, 117903, 127004, 127712; 010-307-1006/G07001030710060/ 102707, 106202, 118004, 129302; 010-307-1006-2/G070010307100620/ 117512, 127718, 114304; 010-307-1208/G07001030712080/ 102709, 105601, 117504; 010-307-1410/G07001030714100/ 102710, 105602, 127713; 010-307-1612/G07001030716120/ 103012, 105603, 117505; 010-313-1205/G07001031312050/ 103013, 105604, 117506, 129303; 010-313-1407/G07001031314070/ 102711, 105605, 117507, 124107, 128702, 130102; 010-313-1609/G07001031316090/ 103014, 105606; 010-313-1811/G07001031318110/ 103015, 105607, 127714; 010-313-2013/G07001031320130/ 103016, 105608, 127715; 010-319-1606/G07001031916060/ 102712, 105609, 114905, 117508; 010-319-1808/G07001031918080/ 103017, 105610, 125103; 010-319-2010/G07001031920100/ 103018, 105611, 125104; 010-325-2007/G07001032520070/ 102713, 105612, 121206; 010-325-2209/G07001032522090/ 105613, 121207, 101206, 101207, 101208
Quantity recalled1259

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2213-2026
FDA event ID
98835
Recalling firm
Novapproach Spine, Alachua, FL
Classification
Class II
Status
Ongoing
Recall initiated
23 Apr 2026
FDA classified
20 May 2026
Reported
27 May 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Lead & Heavy Metals
Device type
Spinal Implants

Timeline

23 Apr 2026Recall initiated by company
20 May 2026Classified by FDA+27 days
27 May 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.

Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.

Product as listed by the FDA

OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,

Where it was distributed

US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.

Nationwide CaliforniaConnecticutFloridaMichiganNew YorkTexas

Questions about this recall

Is OneLIF Interbody Cage ( recalled?

Yes. Novapproach Spine recalled OneLIF Interbody Cage ( on 23 Apr 2026. The recall is Class II and its status is Ongoing. Recall number Z-2213-2026.

Why was it recalled?

Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.

Which lots are affected?

REF/UDI-DI/ Cage Lots: 010-107-1007/G07001010710070/ 103003, 105701, 118001, 129301; 010-107-1007-2/G070010107100720/ 114301, 117509, 128703; 010-107-1209/G07001010712090/ 102714, 105702, 117501; 010-107-1411/G07001010714110/ 103019, 105703; 010-107-1613/G07001010716130/ 103020, 105704, 127701; 010-113-1006/G07001011310060/ 103021, 105705, 118002, 129801; 010-113-1006-2/G070010113100620/ 117510, 127716, 114302; 010-113-1208/G07001011312080/ 103022, 105706, 117502, 126902; 010-113-1410/G07001011314100/ 102715, 105707, 127702; 010-113-1612/G07001011316120/ 103023, 105708, 127703; 010-119-1407/G07001011914070/ 103024, 105709, 125101, 128701; 010-119-1609/G07001011916090/ 102716, 105710, 127001; 010-119-1811/G07001011918110/ 103025, 105711, 127003; 010-125-1606/G07001012516060/ 102717, 117901, 127002; 010-125-1808/G07001012518080/ 103005, 117902, 127704; 010-207-1007/G07001020710070/ 102702, 106201, 118003, 125102; 010-207-1007-2/G070010207100720/ 117511, 124106, 127717, 114303; 010-207-1209/G07001020712090/ 102701, 105401, 124101, 127705, 101201; 010-207-1411/G07001020714110/ 105001, 105402, 101203; 010-207-1613/G07001020716130/ 103007, 105403, 124102, 127706; 010-213-1206/G07001021312060/ 105002, 105404, 121202, 124103, 127707, 101204; 010-213-1408/G07001021314080/ 102704, 105405, 121203, 124104, 127708; 010-213-1610/G07001021316100/ 103008, 105406, 130101; 010-213-1812/G07001021318120/ 103009, 105407, 127709; 010-219-1406/G07001021914060/ 102705, 105408, 117503, 119301, 121201, 124105; 010-219-1608/G07001021916080/ 103010, 105409, 121204, 127710; 010-219-1810/G07001021918100/ 103011, 105410, 127711; 010-225-1807/G07001022518070/ 102706, 113701, 117001, 121205; 010-225-2009/G07001022520090/ 103001, 117903, 127004, 127712; 010-307-1006/G07001030710060/ 102707, 106202, 118004, 129302; 010-307-1006-2/G070010307100620/ 117512, 127718, 114304; 010-307-1208/G07001030712080/ 102709, 105601, 117504; 010-307-1410/G07001030714100/ 102710, 105602, 127713; 010-307-1612/G07001030716120/ 103012, 105603, 117505; 010-313-1205/G07001031312050/ 103013, 105604, 117506, 129303; 010-313-1407/G07001031314070/ 102711, 105605, 117507, 124107, 128702, 130102; 010-313-1609/G07001031316090/ 103014, 105606; 010-313-1811/G07001031318110/ 103015, 105607, 127714; 010-313-2013/G07001031320130/ 103016, 105608, 127715; 010-319-1606/G07001031916060/ 102712, 105609, 114905, 117508; 010-319-1808/G07001031918080/ 103017, 105610, 125103; 010-319-2010/G07001031920100/ 103018, 105611, 125104; 010-325-2007/G07001032520070/ 102713, 105612, 121206; 010-325-2209/G07001032522090/ 105613, 121207, 101206, 101207, 101208

Where was it sold?

US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Novapproach Spine

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIOneLIF Torque Limiting Adapter, AO to 1/4 SQ, 25Nm Limit recalledNovapproach SpineDevice MalfunctionNationwide

Other Spinal Implants recalls