CLASS II ONGOING Device recall · Reported 5 Jun 2024 · FDA Enforcement Report

Ophthalmic Knife Component: N/A recalled by Alcon Research

Alcon Research is recalling Ophthalmic Knife Component: N/A, distributed nationwide in the US. The recall was initiated on 22 Apr 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel/Catalog Number: 8065771540; UDI/DI: 00380657715404; Lot/Batch: 1668D7, 167PRA, 167PTK, 168NMF, 169KL4, 169KTP, 16ALPE, 16ALTU, 16AV18, 16DDPR, 16DDRU;
Quantity recalled19,867 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1924-2024
FDA event ID
94501
Recalling firm
Alcon Research, Fort Worth, TX
Classification
Class II
Status
Ongoing
Recall initiated
22 Apr 2024
FDA classified
30 May 2024
Reported
5 Jun 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

22 Apr 2024Recall initiated by company
30 May 2024Classified by FDA+38 days
5 Jun 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Corneal/Scleral V-Lance Ophthalmic knives.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

ClearCut S Safety Sideport Knife 1.0mm Dual Bevel Model/Catalog Number: 8065771540 Software Version: N/A Product Description: Ophthalmic Knife Component: N/A

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Korea, Kuwait, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom. Product distributed as sterile stand-alone knives have been placed on HOLD. Product included within Alcon Custom Pak (surgical procedure pack) are continuing to be distributed as ceasing distribution would cause immediate market shortage and result in canceled surgeries.

Nationwide Puerto Rico

Questions about this recall

Is Ophthalmic Knife Component: N/A recalled?

Yes. Alcon Research recalled Ophthalmic Knife Component: N/A on 22 Apr 2024. The recall is Class II and its status is Ongoing. Recall number Z-1924-2024.

Why was it recalled?

Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Corneal/Scleral V-Lance Ophthalmic knives.

Which lots are affected?

Model/Catalog Number: 8065771540; UDI/DI: 00380657715404; Lot/Batch: 1668D7, 167PRA, 167PTK, 168NMF, 169KL4, 169KTP, 16ALPE, 16ALTU, 16AV18, 16DDPR, 16DDRU;

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Columbia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Korea, Kuwait, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom. Product distributed as sterile stand-alone knives have been placed on HOLD. Product included within Alcon Custom Pak (surgical procedure pack) are continuing to be distributed as ceasing distribution would cause immediate market shortage and result in canceled surgeries.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Alcon Research

All 107

Other Medical Software recalls

All 903