CLASS II Device recall · Reported 24 Aug 2016 · FDA Enforcement Report
Optical Biometer Al-Scan recalled by Nidek
Nidek is recalling Optical Biometer Al-Scan, distributed nationwide in the US. The recall was initiated on 1 Jul 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2587-2016
- FDA event ID
- 74595
- Recalling firm
- Nidek, Fremont, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 1 Jul 2016
- FDA classified
- 17 Aug 2016
- Reported
- 24 Aug 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Contact Lenses & Eye Devices
Timeline
| 1 Jul 2016 | Recall initiated by company | |
| 17 Aug 2016 | Classified by FDA | +47 days |
| 24 Aug 2016 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Nidek Inc. received information from our manufacturer NIDEK CO. Japan that the shown axial lengths were much shorter at AL-Scan measurement.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
OPTICAL BIOMETER AL-Scan: Software Version: 1.09 and earlier. Ophthalmic: The OPTICAL BIOMETER AL-Scan is a measuring device that can measure as a single unit the values necessary to calculate the power of an IOL for cataract surgery such as axial length, corneal curvature radius, and anterior chamber depth. It measures these necessary values successively through a non-contact optical measurement method.
Where it was distributed
US Nationwide Distribution in the states of: NY, PA. GA. AZ, VA, DE, ME,. MO, TN, and NM.
Nationwide ArizonaDelawareGeorgiaMaineMissouriNew MexicoNew YorkPennsylvaniaTennesseeVirginia
Questions about this recall
Is Optical Biometer Al-Scan recalled?
Yes. Nidek recalled Optical Biometer Al-Scan on 1 Jul 2016. The recall is Class II and its status is Terminated. Recall number Z-2587-2016.
Why was it recalled?
Nidek Inc. received information from our manufacturer NIDEK CO. Japan that the shown axial lengths were much shorter at AL-Scan measurement.
Which lots are affected?
OPTICAL BIOMETER AL-Scan: Software Version: 1.09 and earlier. Serial numbers: 220005, 220006, 220008, 220009, 220013, 220017, 220018, 220019, 220020, 220021, 220023, 220025, 220026, 220031, 220032, 220033, 220036.
Where was it sold?
US Nationwide Distribution in the states of: NY, PA. GA. AZ, VA, DE, ME,. MO, TN, and NM.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Nidek
All 21| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Specular Microscope recalled by Nidek | Nidek | Other | Nationwide |
| Device | CLASS III | RT-5100 Refractor, a component of the Epic-5100 System recalled by Nidek | Nidek | Other | FL |
| Device | CLASS II | Auto REF/Keratometer Ark-1s recalled by Nidek | Nidek | Other | 26 states |
| Device | CLASS II | Final Fit recalled by Nidek | Nidek | Device Malfunction | Nationwide |
| Device | CLASS II | Specular Microscope Cem recalled by Nidek | Nidek | Device Malfunction | 18 states |
Other Contact Lenses & Eye Devices recalls
All 152| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Tecnis Eyhance Iol with Tecnis Simplicity" Delivery System recalled | AMO Puerto Rico Manufacturing | Sterility | Nationwide |
| Device | CLASS II | Alcon PRECISION1 One-Day Contact Lenses, -2.25D recalled | Alcon Research | Sterility | 19 states |
| Device | CLASS II | Sofmed Breathables, (stenfilcon A) contact lens recalled by CooperVision | CooperVision | Mold | Nationwide |
| Device | CLASS II | Voyant 1-Day Premium Toric, (stenfilcon A) contact lens recalled | CooperVision | Mold | Nationwide |
| Device | CLASS II | MyDay Toric, stenfilcon A) Contact Lens recalled by CooperVision | CooperVision | Mold | Nationwide |