CLASS II ONGOING Device recall · Reported 4 Nov 2026 · FDA Enforcement Report

Orthoscan Versa Accessory Cart recalled by Ziehm-orthoscan

Ziehm-orthoscan is recalling Orthoscan Versa Accessory Cart, distributed nationwide in the US. The recall was initiated on 3 Sep 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI/ Mini C-Arm serial number (6Axxxx)/Accessory cart serial Number: (VCXXX): 1 6A0110 /VC0110 2 6A0111 /VC0111 3 6A0112 /VC0112 4 6A0115 /VC0113 5 6A0114 /VC0114 6 6A0117 /VC0115 7 6A0116 /VC0116 8 6A0113 /VC0117 9 6A0118 /VC0118 10 6A0119/ VC0119, VC0120; 11 6A0120 /VC0121; 12 6A0121 /VC0122; 13 6A0123 /VC0123; 14 6A0122 /VC0124; 15 6A0126 /VC0125; 16 6A0133 /VC0134; 17 6A0124 /VC0126; 18 6A0128 /VC0128; 19 6A0127 /VC0129; 20 6A0129 /VC0130; 21 6A0125 /VC0127; 22 6A0131 /VC0132; 23 6A0134 /VC0133; 24 6A0142 /VC0144; 25 6A0136 /VC0136; 26 6A0132 /VC0137; 27 6A0137 /VC0138; 28 6A0138 /VC0139; 29 6A0139 /VC0140; 30 6A0140 /VC0141; 31 6A0141 /VC0142; 32 6A0146 /VC0147; 33 6A0143 /VC0143; 34 6A0144 /VC0145; 35 6A0152 /VC0153; 36 6A0147 /VC0148; 37 6A0148 /VC0149; 38 6A0149 /VC0150; 39 6A0150 /VC0151; 40 6A0151 / VC0152; 41 6A0163 / VC0165; 42 6A0130 / VC0131; 43 6A0155 / VC0156; 44 6A0157 / VC0157; 45 6A0156/ VC0158; 46 6A0158 / VC0159, VC0160; 47 6A0159/ VC0161; 48 6A0160/ VC0162; 49 6A0161 / VC0163; 50 6A0162 VC0164; 51 6A0153 VC0154; 52 6A0164 VC0166; 53 6A0165 VC0167
Quantity recalled55 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0286-2027
FDA event ID
99979
Recalling firm
Ziehm-orthoscan, Scottsdale, AZ
Classification
Class II
Status
Ongoing
Recall initiated
3 Sep 2026
FDA classified
23 Oct 2026
Reported
4 Nov 2026
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

3 Sep 2026Recall initiated by company
23 Oct 2026Classified by FDA+50 days
4 Nov 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020

Where it was distributed

Worldwide - US Nationwide distribution in the states of AZ , CA , CO , CT , FL , GA , IL , IN , KS, MA , MD , MI , MN , NC , NE , NV , NY , OH , OR , SC, TX , UT , WA and the countries of Germany , Australia , Spain , Switzerland, Japan.

Nationwide ArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIllinoisIndianaKansasMassachusettsMarylandMichiganMinnesotaNorth CarolinaNebraskaNevadaNew YorkOhioOregonSouth Carolina

Questions about this recall

Why was it recalled?

The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.

Which lots are affected?

UDI/ Mini C-Arm serial number (6Axxxx)/Accessory cart serial Number: (VCXXX): 1 6A0110 /VC0110 2 6A0111 /VC0111 3 6A0112 /VC0112 4 6A0115 /VC0113 5 6A0114 /VC0114 6 6A0117 /VC0115 7 6A0116 /VC0116 8 6A0113 /VC0117 9 6A0118 /VC0118 10 6A0119/ VC0119, VC0120; 11 6A0120 /VC0121; 12 6A0121 /VC0122; 13 6A0123 /VC0123; 14 6A0122 /VC0124; 15 6A0126 /VC0125; 16 6A0133 /VC0134; 17 6A0124 /VC0126; 18 6A0128 /VC0128; 19 6A0127 /VC0129; 20 6A0129 /VC0130; 21 6A0125 /VC0127; 22 6A0131 /VC0132; 23 6A0134 /VC0133; 24 6A0142 /VC0144; 25 6A0136 /VC0136; 26 6A0132 /VC0137; 27 6A0137 /VC0138; 28 6A0138 /VC0139; 29 6A0139 /VC0140; 30 6A0140 /VC0141; 31 6A0141 /VC0142; 32 6A0146 /VC0147; 33 6A0143 /VC0143; 34 6A0144 /VC0145; 35 6A0152 /VC0153; 36 6A0147 /VC0148; 37 6A0148 /VC0149; 38 6A0149 /VC0150; 39 6A0150 /VC0151; 40 6A0151 / VC0152; 41 6A0163 / VC0165; 42 6A0130 / VC0131; 43 6A0155 / VC0156; 44 6A0157 / VC0157; 45 6A0156/ VC0158; 46 6A0158 / VC0159, VC0160; 47 6A0159/ VC0161; 48 6A0160/ VC0162; 49 6A0161 / VC0163; 50 6A0162 VC0164; 51 6A0153 VC0154; 52 6A0164 VC0166; 53 6A0165 VC0167

Where was it sold?

Worldwide - US Nationwide distribution in the states of AZ , CA , CO , CT , FL , GA , IL , IN , KS, MA , MD , MI , MN , NC , NE , NV , NY , OH , OR , SC, TX , UT , WA and the countries of Germany , Australia , Spain , Switzerland, Japan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 10 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other Other Medical Devices recalls

All 13,968