CLASS II ONGOING Device recall · Reported 28 Jan 2026 · FDA Enforcement Report

Patient Information Center iX recalled by Philips North America

Philips North America is recalling Patient Information Center iX, distributed nationwide in the US. The recall was initiated on 22 Dec 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Numbers: (1) 866389, (2) 866390; UDI-DI: (1) 00884838127074, (2) N/A; Serial Numbers: 4Z24-5CKL-P, 0P6C-4TL3-D, 6J58-3LL6-0, 216G-7PGV-E, 0F2U-1BLC-H, 291E-7HJU-A, 1M6W-1UP5-1, 4620-0XNW-5, 4W5R-3FHA-L, 6B73-3ZKP-K, 3E79-49M4-3, 4F6Y-70HU-P, 0128-2WG5-0, 0830-3RLW-W, 097K-7LPC-5, 0D2V-1BLC-8, 120T-17M9-A, 173C-20JM-J, 190B-64MU-U, 1K2J-15LB-P, 1L3T-7KGB-E, 2K06-4ZPR-X, 3916-5DJ6-Z, 3V7H-06MT-E, 412A-2WG5-Y, 4271-2NL7-M, 4Z3L-5PP3-0, 5D0E-3EH3-3, 5E0B-3EH3-D, 5E25-14LG-9, 657X-31N5-L, 6G5U-3FHA-R, 781T-7TJA-B, 7N3U-51MJ-B, 7P31-7NPM-P, 283X-7ZPB-A, 7H77-09GU-Z, 4C59-26K6-P, 4N0L-0FKE-2, 5Y0A-02P6-L, 494H-1CJJ-3, 326G-7PGV-A, 292G-3YNZ-9, 0V2A-6DHP-0, 6354-58M8-C, 4W5B-6JK9-A, 5M7K-0VKT-T, 3401-1FHE-M, 6R30-5BNT-N, 6N2E-7YK1-1, 6R0H-6AMX-T, 107C-62PF-Y, 2X1T-7TJA-H, 1M6G-7PGV-K, 1C4Z-6KKD-H, 0B3Z-2RHR-L, 7H39-3JGK-U, 0M0V-6FKB-5, 793D-41PZ-X, 233G-46MR-V, 782Z-3RLW-D, 5N2B-5YJP-G, 0L5F-0ZNF-J, 4G4N-2LKY-K, 2M48-4XM7-Y, 133W-3UN7-B, 295N-0YG8-Y, 5D7W-00G0-1, 5R78-6CPV-0, 3C79-49M4-9, 3T24-5CKL-A, 6H74-27LY-T, 186W-1UP5-8, 0021-1DNN-P, 0247-7ELE-R, 0D5Y-32PN-A, 0J7W-31N5-2, 0R2W-1VNP-P, 0Y5Z-7FHX-N, 121L-1KG1-M, 1366-7XLV-X, 144W-7RPJ-5, 1472-5WKF-X, 1C19-1MHK-F, 1G0F-69LP-G, 1J5U-5HMV-N, 1P4F-34MB-W, 1P5C-6JK9-D, 212R-6HHJ-P, 221V-7TJA-4, 227N-7JG6-J, 232W-7DL9-G, 2A5X-2PJT-0, 2A75-27LY-3, 2F5J-0JLM-7, 2J19-1MHK-W, 2L16-59N8-M, 2M73-5WKF-U, 2N3V-6YM2-X, 2P6J-7PGV-E, 2Y5M-5AGH-E, 326W-21KC-B, 332M-47P7-Y, 3A29-0CKW-3, 3C0U-7AHH-6, 3C2U-05HF-W, 3E05-4ZPR-U, 3G6H-39G4-Y, 3H1T-22HW-5, 3H5R-5HMV-M, 3H79-09GU-5, 3J2K-2TP9-1, 3W1Y-3CH6-J, 4L7Y-00G0-F, 4M77-1HM0-G, 4N3D-20JM-K, 4R0G-60JV-1, 4V2Z-7NPM-M, 4W08-0UH7-U, 4X6C-1PJ5-J, 4Y08-0UH7-L, 500W-6WNX-A, 594D-0EG2-9, 5U0A-3WPK-C, 5W1G-50P2-N, 5Z6J-3HL1-B, 617W-31N5-F, 637V-7BG9-W, 673T-67ME-B, 686N-3NNV-6, 6D3P-4WMG-A, 6K70-3XNY-W, 6P42-35J4-V, 6U1T-75JN-Y, 771A-2JPY-0, 7F5L-73HZ-3, 7G3A-13JG-L, 7K0P-4AH2-6, 7Z2R-05HF-X, 1J1P-6PKN-T, 4R7H-7LPC-J, 282N-0WPA-X, 5G35-04K8-E, 366L-2XHV-A, 5L7D-2UGW-T, 4N0L-0FKE-2;
Quantity recalled146 units (41 US, 105 OUS)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1156-2026
FDA event ID
98217
Recalling firm
Philips North America, Cambridge, MA
Classification
Class II
Status
Ongoing
Recall initiated
22 Dec 2025
FDA classified
22 Jan 2026
Reported
28 Jan 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

22 Dec 2025Recall initiated by company
22 Jan 2026Classified by FDA+31 days
28 Jan 2026Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Patient Information Center iX; Software Version Number: 4.5.0

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Canada, Czech Republic, Dominican Rep, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Netherlands, Norway, Poland, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Utd.Arab.Emir.

Nationwide

Questions about this recall

Why was it recalled?

A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.

Which lots are affected?

Model Numbers: (1) 866389, (2) 866390; UDI-DI: (1) 00884838127074, (2) N/A; Serial Numbers: 4Z24-5CKL-P, 0P6C-4TL3-D, 6J58-3LL6-0, 216G-7PGV-E, 0F2U-1BLC-H, 291E-7HJU-A, 1M6W-1UP5-1, 4620-0XNW-5, 4W5R-3FHA-L, 6B73-3ZKP-K, 3E79-49M4-3, 4F6Y-70HU-P, 0128-2WG5-0, 0830-3RLW-W, 097K-7LPC-5, 0D2V-1BLC-8, 120T-17M9-A, 173C-20JM-J, 190B-64MU-U, 1K2J-15LB-P, 1L3T-7KGB-E, 2K06-4ZPR-X, 3916-5DJ6-Z, 3V7H-06MT-E, 412A-2WG5-Y, 4271-2NL7-M, 4Z3L-5PP3-0, 5D0E-3EH3-3, 5E0B-3EH3-D, 5E25-14LG-9, 657X-31N5-L, 6G5U-3FHA-R, 781T-7TJA-B, 7N3U-51MJ-B, 7P31-7NPM-P, 283X-7ZPB-A, 7H77-09GU-Z, 4C59-26K6-P, 4N0L-0FKE-2, 5Y0A-02P6-L, 494H-1CJJ-3, 326G-7PGV-A, 292G-3YNZ-9, 0V2A-6DHP-0, 6354-58M8-C, 4W5B-6JK9-A, 5M7K-0VKT-T, 3401-1FHE-M, 6R30-5BNT-N, 6N2E-7YK1-1, 6R0H-6AMX-T, 107C-62PF-Y, 2X1T-7TJA-H, 1M6G-7PGV-K, 1C4Z-6KKD-H, 0B3Z-2RHR-L, 7H39-3JGK-U, 0M0V-6FKB-5, 793D-41PZ-X, 233G-46MR-V, 782Z-3RLW-D, 5N2B-5YJP-G, 0L5F-0ZNF-J, 4G4N-2LKY-K, 2M48-4XM7-Y, 133W-3UN7-B, 295N-0YG8-Y, 5D7W-00G0-1, 5R78-6CPV-0, 3C79-49M4-9, 3T24-5CKL-A, 6H74-27LY-T, 186W-1UP5-8, 0021-1DNN-P, 0247-7ELE-R, 0D5Y-32PN-A, 0J7W-31N5-2, 0R2W-1VNP-P, 0Y5Z-7FHX-N, 121L-1KG1-M, 1366-7XLV-X, 144W-7RPJ-5, 1472-5WKF-X, 1C19-1MHK-F, 1G0F-69LP-G, 1J5U-5HMV-N, 1P4F-34MB-W, 1P5C-6JK9-D, 212R-6HHJ-P, 221V-7TJA-4, 227N-7JG6-J, 232W-7DL9-G, 2A5X-2PJT-0, 2A75-27LY-3, 2F5J-0JLM-7, 2J19-1MHK-W, 2L16-59N8-M, 2M73-5WKF-U, 2N3V-6YM2-X, 2P6J-7PGV-E, 2Y5M-5AGH-E, 326W-21KC-B, 332M-47P7-Y, 3A29-0CKW-3, 3C0U-7AHH-6, 3C2U-05HF-W, 3E05-4ZPR-U, 3G6H-39G4-Y, 3H1T-22HW-5, 3H5R-5HMV-M, 3H79-09GU-5, 3J2K-2TP9-1, 3W1Y-3CH6-J, 4L7Y-00G0-F, 4M77-1HM0-G, 4N3D-20JM-K, 4R0G-60JV-1, 4V2Z-7NPM-M, 4W08-0UH7-U, 4X6C-1PJ5-J, 4Y08-0UH7-L, 500W-6WNX-A, 594D-0EG2-9, 5U0A-3WPK-C, 5W1G-50P2-N, 5Z6J-3HL1-B, 617W-31N5-F, 637V-7BG9-W, 673T-67ME-B, 686N-3NNV-6, 6D3P-4WMG-A, 6K70-3XNY-W, 6P42-35J4-V, 6U1T-75JN-Y, 771A-2JPY-0, 7F5L-73HZ-3, 7G3A-13JG-L, 7K0P-4AH2-6, 7Z2R-05HF-X, 1J1P-6PKN-T, 4R7H-7LPC-J, 282N-0WPA-X, 5G35-04K8-E, 366L-2XHV-A, 5L7D-2UGW-T, 4N0L-0FKE-2;

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Canada, Czech Republic, Dominican Rep, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Netherlands, Norway, Poland, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Utd.Arab.Emir.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Philips North America

All 558

Other Medical Software recalls

All 932