CLASS II Device recall · Reported 29 Aug 2012 · FDA Enforcement Report
Positron Emission Tomography (PET) recalled by Philips Medical Systems
Philips Medical Systems (Cleveland) is recalling Positron Emission Tomography (PET) and X-ray Computed Tomography, Gemini TF 16 PET/CT, distributed nationwide in the US. The recall was initiated on 8 Aug 2012 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2239-2012
- FDA event ID
- 62892
- Recalling firm
- Philips Medical Systems (Cleveland), Cleveland, OH
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 8 Aug 2012
- FDA classified
- 22 Aug 2012
- Reported
- 29 Aug 2012
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Heart & Vascular Devices
Timeline
| 8 Aug 2012 | Recall initiated by company | |
| 22 Aug 2012 | Classified by FDA | +14 days |
| 29 Aug 2012 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed within the UPS battery cabinet, entered thermal runaway causing the battery to overheat.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Positron Emission Tomography (PET) and X-ray Computed Tomography, GEMINI TF 16 PET/CT, GEMINI GXL 16 PET/CT, GEMINI TF Big Bore PET/CT, Philips Medical Systems, Cleveland, OH. The device is a diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart and other organ imaging). Attenuation correction is accomplished by CTAC. The device also provides for list mode, dynamic, and gated acquisitions. Image processing and display workstations provide software applications to process, analyze, display, quantify and interpret medical images/data. The PET and CT images may be registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined metabolic and anatomical data at different angles. Trained professionals use the images in: " The evaluation, detection and diagnosis of lesions, disease and organ function such as but not limited to cancer, cardiovascular disease, and neurological disorders. " The detection, localization, and staging of tumors and diagnosing cancer patients. " Treatment planning and interventional radiology procedures. The device includes software that provides a quantified analysis of with regional cerebral activity from PET images.
Where it was distributed
Nationwide Distribution-including the states of CO, FL, MD, MN, NE and PA.
Nationwide ColoradoFloridaMarylandMinnesotaNebraskaPennsylvania
Questions about this recall
Why was it recalled?
Philips became aware of an issue relating to a UPS battery cabinet that appeared to have burned a hole through the casing on one of the batteries. Supplier of the UPS, Chloride/Emerson, determined the batteries, housed within the UPS battery cabinet, entered thermal runaway causing the battery to overheat.
Which lots are affected?
GEMINI TF 16 System serial numbers: 7194, 7186, 7235, 7233 GEMINI GXL 16 System serial number: 4201 GEMINI TF Big Bore System serial numbers: 9022, 9201
Where was it sold?
Nationwide Distribution-including the states of CO, FL, MD, MN, NE and PA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips Medical Systems (Cleveland)
All 315| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Pinnacle Radiation Therapy Planning System recalled | Philips Medical Systems (Cleveland) | Device Malfunction | Nationwide |
| Device | CLASS II | Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System recalled | Philips Medical Systems (Cleveland) | Device Malfunction | Nationwide |
| Device | CLASS II | Pinnacle3 Radiation Therapy Planning System recalled | Philips Medical Systems (Cleveland) | Other | 21 states |
| Device | CLASS I | Forte Gamma Camera System is intended to produce images recalled | Philips Medical Systems (Cleveland) | Other | Nationwide |
| Device | CLASS I | Forte Gamma Camera System is intended to produce images recalled | Philips Medical Systems (Cleveland) | Other | Nationwide |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |