CLASS II ONGOING Device recall · Reported 4 May 2022 · FDA Enforcement Report
PrisMax System, Product recalled by Baxter Healthcare
Baxter Healthcare is recalling PrisMax System, Product, distributed nationwide in the US. The recall was initiated on 15 Mar 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0991-2022
- FDA event ID
- 89823
- Recalling firm
- Baxter Healthcare, Deerfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 15 Mar 2022
- FDA classified
- 22 Apr 2022
- Reported
- 4 May 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 15 Mar 2022 | Recall initiated by company | |
| 22 Apr 2022 | Classified by FDA | +38 days |
| 4 May 2022 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
PrisMax System, Product Code 955626
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Mexico, Brazil, Jamaica, Trinidad and Tobago, Canada, Australia, Singapore, Hong Kong, Vietnam, New Zealand, Korea, Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Questions about this recall
Is PrisMax System, Product recalled?
Yes. Baxter Healthcare recalled PrisMax System, Product on 15 Mar 2022. The recall is Class II and its status is Ongoing. Recall number Z-0991-2022.
Why was it recalled?
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
Which lots are affected?
GTIN 00085412639499, All Serial Numbers
Where was it sold?
Worldwide distribution - US Nationwide and the countries of Mexico, Brazil, Jamaica, Trinidad and Tobago, Canada, Australia, Singapore, Hong Kong, Vietnam, New Zealand, Korea, Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Baxter Healthcare
All 562| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Baxter Novum IQ Syringe Pump, Loaner recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Baxter Novum IQ Syringe Pump recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Drug | CLASS II | Clindamycin Phosphate in 9% Sodium Chloride Injection recalledClindamycin Phosphate | Baxter Healthcare | Sterility | Nationwide |
| Drug | CLASS II | Vancomycin Injection, USP in 5% Dextrose per (5 mg/mL) recalledVancomycin Hydrochloride | Baxter Healthcare | Sterility | Nationwide |
| Drug | CLASS II | Clindamycin Injection USP in 5% Dextrose, 900 mg recalledClindamycin Phosphate | Baxter Healthcare | Sterility | Nationwide |
Other Other Medical Devices recalls
All 13,771| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: Blood Draw Kit recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ortho Aspiration Kit recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep DYKA1054F recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Arterial Line Insertion Kit recalled | Medline Industries | Sterility | Nationwide |