CLASS II ONGOING Device recall · Reported 11 Oct 2023 · FDA Enforcement Report
Puritan Bennett Cuff Pressure Manager recalled by Covidien
Covidien is recalling Puritan Bennett Cuff Pressure Manager, distributed in 16 states. The recall was initiated on 22 Aug 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2651-2023
- FDA event ID
- 92941
- Recalling firm
- Covidien, Boulder, CO
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 22 Aug 2023
- FDA classified
- 29 Sep 2023
- Reported
- 11 Oct 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Lead & Heavy Metals
- Device type
- Other Medical Devices
Timeline
| 22 Aug 2023 | Recall initiated by company | |
| 29 Sep 2023 | Classified by FDA | +38 days |
| 11 Oct 2023 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
Puritan Bennett Cuff Pressure Manager, REF: 180-03
Where it was distributed
US: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands
AlabamaArizonaCaliforniaColoradoFloridaIllinoisKansasKentuckyMassachusettsMichiganNorth CarolinaNew JerseyNew YorkTennesseeTexasUtah
Stores named: Walmart
Questions about this recall
Is Puritan Bennett Cuff Pressure Manager recalled?
Yes. Covidien recalled Puritan Bennett Cuff Pressure Manager on 22 Aug 2023. The recall is Class II and its status is Ongoing. Recall number Z-2651-2023.
Why was it recalled?
Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.
Which lots are affected?
UDI-DI: 10884521782495, Serial Numbers: E223500003, E223500004, E223500005, E223500006, E223500007, E223500008, E223500009, E223500010, E223500011, E223500012, E223500013, E223500014, E223500015, E223500016, E223500017, E223500018, E223500019, E223500020, E223500021, E223500022, E223500023, E223500024, E223500025, E223500026, E223500027, E223500028, E223500029, E223500030, E223500031, E223500032, E223500054, E223500055, E223500079, E223500080, E223500081, E223500082, E223500083, E223600009, E223600010, E223600011, E223600012, E223600013, E223600014, E223600015, E223600016, E223600017, E223600018, E223600019, E223600020, E223600021, E223600022, E223600023, E223600024, E223600042, E223600070, E223600071, E223600072, E223600073, E223600074, E223600075, E223600076, E223600077, E223600078, E223600079, E223600080, E223600081, E223600082, E223600083, E223600084, E223600085, E223600086, E223600087, E223600088, E223600089, E223600090, E223600092, E223600093, E223600095, E223600096, E223600097, E223600098, E223600099, E223600100, E224900001, E224900002, E224900003, E224900004, E224900005, E224900006, E224900007, E224900008, E224900009, E224900010, E224900011, E224900012, E224900013, E224900014, E224900015, E224900017, E224900018, E224900019, E224900020, E224900021, E224900022, E224900023, E224900024, E224900025, E224900026, E224900027, E224900028, E224900029, E224900030, E224900031, E224900032, E224900033, E224900034, E224900035, E224900036, E224900037, E224900039, E224900040, E224900041, E224900042, E224900043, E224900044, E224900045, E224900047, E224900048, E224900049, E224900050, E224900052, E224900053, E224900055, E224900056, E224900057, E224900058, E224900059, E224900060, E224900061, E224900062, E224900064, E224900065, E224900066, E224900067, E224900068, E224900069, E224900070, E224900072, E224900073, E224900074, E224900075, E224900076, E224900077, E224900078, E224900081, E224900082, E224900083, E224900084, E224900085, E224900087, E224900091, E224900092, E224900093, E224900094, E224900095, E224900098, E224900099, E224900100, E225000001, E225000002, E225000003, E225000004, E225000005, E225000006, E225000007, E225000011, E225000012, E225000013, E225000014, E225000015, E225000016, E225000017, E225000018, E225000019, E225000020, E225000021, E225000022, E225000023, E225000024, E225000025, E225000026, E225000027, E225000029, E225000031, E225000032, E225000033, E225000034, E225000035, E225000036, E225000037, E225000038, E225000039, E225000040, E225000041, E225000042, E225000043, E225000044, E225000045, E225000046, E225000047, E225000048, E225000049, E225000050, E225000051, E225000052, E225000053, E225000054, E225000055, E225000056, E225000057, E225000058, E225000059, E225000060, E225000061, E225000062, E225000063, E225000064, E225000065, E225000066, E225000067, E225000068, E225000069, E225000070, E225000071, E225000072, E225000073, E225000075, E225000076, E225000077, E225000078, E225000079, E225000080, E225000081, E225000082, E225000083, E225000084, E225000085, E225000086, E225000087, E225000088, E225000089, E225000090, E225000091, E225000092, E225000093, E225000094, E225000095, E225000096, E225000097, E225000098, E225000099, E225000100, E230200001, E230200002, E230200003, E230200004, E230200005, E230200006, E230200007, E230200008, E230200009, E230200010, E230200011, E230200012, E230200013, E230200014, E230200015, E230200016, E230200017, E230200018, E230200019, E230200020, E230200021, E230200022, E230200023, E230200024, E230200025, E230200027, E230200028, E230200029, E230200030, E230200031, E230200032, E230200033, E230200034, E230200035, E230200036, E230200037, E230200038, E230200039, E230200040, E230200041, E230200042, E230200043, E230200044, E230200045, E230200046, E230200048, E230200049, E230200050, E230200051, E230200052, E230200053, E230200054, E230200056, E230200057, E230200058, E230200059, E230200060, E230200061, E230200062, E230200064, E230200065, E230200066, E230200067, E230200068, E230200069, E230200070, E230200071, E230200072, E230200073, E230200074, E230200075, E230200076, E230200077, E230200078, E230200079, E230200080, E230200081, E230200082, E230200083, E230200085, E230200086, E230200087, E230200088, E230200089, E230200090, E230200091, E230200092, E230200095, E230200096, E230500006, E230500007, E230500008, E230500009, E230500010, E230500011, E230500031, E230500032, E230500033, E230500034, E230500035, E230500036, E230500037, E230500038, E230500039, E230500040, E230500041, E230500042, E230500043, E230500044, E230500045, E230500046, E230500047, E230500048, E230500049, E230500050, E230500051, E230500052, E230500053, E230500054, E230500055, E230500056, E230500057, E230500058, E230500059, E230500060, E230500061, E230500062, E230500063, E230500064, E230500065, E230500066, E230500092, E230500093, E230500094, E230500103, E230500104, E230500105, E230500106, E230500107, E230500108, E230500109, E230500110, E230500111, E230500112, E230900079, E230900080
Where was it sold?
US: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El Salvador, Sweden, Canary Islands
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Covidien
All 493| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | EndoFLIP System that includes recalled by Covidien | Covidien | Other | Nationwide |
| Device | CLASS II | Dermalon 88861757-31 Dermalon* 4-0 Blu C13 recalled by Covidien | Covidien | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Surgipro SP-623 Surgipro* 2-0 BLU SC2 X36 SP recalled by Covidien | Covidien | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | V-Loc 90 Absorbable Wound Closure Device Vloc 90 0 VL recalled | Covidien | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Surgipro II S8698GX Surgipro II BL 5/0 18 P-13 S8698GX recalled | Covidien | Lead & Heavy Metals | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |