CLASS I Device recall · Reported 5 Sep 2012 · FDA Enforcement Report

On-q pump with Ondemand bolus button, On-q C-bloc with Ondemand recalled

I-Flow is recalling On-q pump with Ondemand bolus button, On-q C-bloc with Ondemand. The recall was initiated on 8 May 2012 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesProducts having a 10-digit Lot # that is greater than 0200521454 (i.e. 0200521455 and greater) are not impacted by this recall.
Quantity recalled80,646 units total (65,926 units in US)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2253-2012
FDA event ID
62769
Recalling firm
I-Flow, Lake Forest, CA
Classification
Class I
Status
Terminated
Recall initiated
8 May 2012
FDA classified
27 Aug 2012
Reported
5 Sep 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

8 May 2012Recall initiated by company
27 Aug 2012Classified by FDA+111 days
5 Sep 2012Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

It was determined that in a small quantity of ON-Q pumps with ONDEMAND, the bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowermost position. As a result, the patient may receive continuous infusion at a rate greater than expected.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

ON-Q pump with ONDEMAND bolus button, ON-Q C-bloc with ONDEMAND; ON-Q C-bloc with Select a Flow and ONDEMAND Model/Reference #: 7000180, 7000181, 7000182, 7000367, 7000397, CB003, CB005, CB006, P400X1-7+5/60, P400X2-14+5/30, P400X5+5/60, and PMB01. Part Numbers: 101347100, 101347300, 101347400, 101360200, 101368400, 101368600, 101368700, 102987000, 102987200, 102987300, 103161800, 103534300, 5001438, 5001470, 5001651, 5001780, 5001808, 5001809, 5001812, 7000180, 7000181, 7000182, & 7000367. Indicated for continuous and or intermittent delivery of medication to or around surgical wound sites and or close proximity to nerves.

Where it was distributed

Worldwide Distribution -- USA, Canada, Australia, the EU (Germany), United Arab Emirates, and Cyprus.

Questions about this recall

Is On-q pump with Ondemand bolus button, On-q C-bloc with Ondemand recalled?

Yes. I-Flow recalled On-q pump with Ondemand bolus button, On-q C-bloc with Ondemand on 8 May 2012. The recall is Class I and its status is Terminated. Recall number Z-2253-2012.

Why was it recalled?

It was determined that in a small quantity of ON-Q pumps with ONDEMAND, the bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowermost position. As a result, the patient may receive continuous infusion at a rate greater than expected.

Which lots are affected?

Products having a 10-digit Lot # that is greater than 0200521454 (i.e. 0200521455 and greater) are not impacted by this recall.

Where was it sold?

Worldwide Distribution -- USA, Canada, Australia, the EU (Germany), United Arab Emirates, and Cyprus.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.