CLASS I ONGOING Device recall · Reported 8 Jul 2026 · FDA Enforcement Report
Reprocessed Electrophysiology Catheters and Ultrasound Catheters recalled
Medline Industries is recalling Reprocessed Electrophysiology Catheters and Ultrasound Catheters, distributed nationwide in the US. The recall was initiated on 15 May 2026 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2615-2026
- FDA event ID
- 99122
- Recalling firm
- Medline Industries, Northfield, IL
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 15 May 2026
- FDA classified
- 2 Jul 2026
- Reported
- 8 Jul 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Particulate Matter
- Device type
- Heart & Vascular Devices
Timeline
| 15 May 2026 | Recall initiated by company | |
| 2 Jul 2026 | Classified by FDA | +48 days |
| 8 Jul 2026 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Product as listed by the FDA
Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.
Where it was distributed
US Nationwide distribution in the state of TX.
Questions about this recall
Why was it recalled?
Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Which lots are affected?
Lot EP250203 Item no. 401206RH - UDI-DI 10197344043767; Item no. 401305RH - UDI-DI 10197344043637.
Where was it sold?
US Nationwide distribution in the state of TX.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Medline Industries
All 738| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Drape C-section W/Pouch ST 7/CS, Medline Product Number recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: Fetal Surgery Cds-lf recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Drape C-section Fen Clr Screen W, Medline Product Number recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: C-section Pack recalled | Medline Industries | Other | Nationwide |
Other Heart & Vascular Devices recalls
All 6,270| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Impella Connect Optical AIC, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS I | Impella Optical Controller, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS I | Automated Impella Controller, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS II | Direxion Torqueable Microcatheter, Ngmc/Fathom/027/Straight/1RO/130 recalled | Boston Scientific Cork | Device Malfunction | Nationwide |