CLASS II ONGOING Device recall · Reported 17 Jan 2024 · FDA Enforcement Report

Scoot Control R-net, Product Number P015 recalled by mo-Vis BVBA

mo-Vis Bvba is recalling Scoot Control R-net, Product Number P015-61, distributed in Tennessee, Texas. The recall was initiated on 28 Nov 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 05407008320119; Lot Numbers: 3304-3317 3322-3329 3440-3452 3569-3618 3830-3844 4020-4054 4170-4176 4250-4255 4321-4330
Quantity recalled129 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0713-2024
FDA event ID
93556
Recalling firm
mo-Vis Bvba, Nevele
Classification
Class II
Status
Ongoing
Recall initiated
28 Nov 2023
FDA classified
11 Jan 2024
Reported
17 Jan 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

28 Nov 2023Recall initiated by company
11 Jan 2024Classified by FDA+44 days
17 Jan 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the result. An unwanted movement could result in damage to the wheelchair, material damage around the chair and/or injury to the user, attendant or by-standers.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Scoot Control R-net, Product Number P015-61. The Scoot Control is intended to support a wheelchair attendant in controlling or maneuvering a power chair

Where it was distributed

US States: TN, TX.

TennesseeTexas

Questions about this recall

Why was it recalled?

In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the result. An unwanted movement could result in damage to the wheelchair, material damage around the chair and/or injury to the user, attendant or by-standers.

Which lots are affected?

UDI-DI: 05407008320119; Lot Numbers: 3304-3317 3322-3329 3440-3452 3569-3618 3830-3844 4020-4054 4170-4176 4250-4255 4321-4330

Where was it sold?

US States: TN, TX.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.