CLASS II ONGOING Device recall · Reported 17 Jun 2026 · FDA Enforcement Report
SD LTM 64 PLUS, EEG Amplifier/recorder recalled by Micromed S.p.A
Micromed is recalling SD LTM 64 PLUS, EEG Amplifier/recorder, distributed nationwide in the US. The recall was initiated on 4 May 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2353-2026
- FDA event ID
- 98924
- Recalling firm
- Micromed, Mogliano Veneto, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 4 May 2026
- FDA classified
- 5 Jun 2026
- Reported
- 17 Jun 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 4 May 2026 | Recall initiated by company | |
| 5 Jun 2026 | Classified by FDA | +32 days |
| 17 Jun 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02
Where it was distributed
Worldwide - US Nationwide distribution in the states of FL, PA.
Questions about this recall
Why was it recalled?
Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
Which lots are affected?
UDI-DI: 8033928120897; Lot Code: Model No SD LTM 64 PLUS; UDI-DI 8033928120897; Serial Numbers BAA-0051/01-19, BAA-0069/02-19, BAA-0070/02-19, BAA-0071/02-19, BAA-0082/02-19, BAA-0083/02-19, BAA-0084/02-19, BAA-0085/02-19, BAA-0086/02-19, BAA-0087/02-19, BAA-0088/02-19, BAA-0089/02-19, BAA-0090/02-19, BAA-0091/02-19, BAA-0092/02-19, BAA-0093/02-19, BAA-0095/02-19, BAA-0096/02-19, BAA-0098/02-19, BAA-0104/02-19, BAA-0105/02-19, BAA-0106/02-19, BAA-0107/02-19, BAA-0108/02-19, BAA-0109/02-19, BAA-0110/02-19, BAA-0111/02-19, BAA-0112/02-19, BAA-0113/02-19, BAA-0115/02-19, BAA-0116/02-19, BAA-0117/02-19, BAA-0118/02-19, BAA-0119/02-19, BAA-0121/02-19, BAA-0122/02-19, BAA-0142/03-19, BAA-0155/03-20, BAA-0156/03-20, BAA-0191/04-20, BAA-0192/04-20, BAA-0193/04-20, BAA-0194/04-20, BAA-0286/04-21, BAA-0287/04-21, BAA-0288/04-21, BAA-0289/04-21, BAA-0329/04-22
Where was it sold?
Worldwide - US Nationwide distribution in the states of FL, PA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other Medical Software recalls
All 932| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Kit: Urinary Clean Catch 20/cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Device Malfunction | CA |
| Device | CLASS II | Spectral CT and Spectral CT 7500 RT recalled by Philips Medical Systems | Philips Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | SmartPath to dStream for XR and 3.0T recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Upgrade to MR recalled by Philips North America | Philips North America | Device Malfunction | Nationwide |
| Device | CLASS II | Positron Emission Tomography and Computed Tomography System recalled | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |