CLASS II Device recall · Reported 29 Jan 2014 · FDA Enforcement Report

Siemens CentraLink" Data Management System V14.0.4 recalled

Siemens Healthcare Diagnostics is recalling Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink, distributed in Minnesota, New Jersey, Nevada. The recall was initiated on 30 Oct 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 Catalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877, 10815474.
Quantity recalled40

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0823-2014
FDA event ID
66694
Recalling firm
Siemens Healthcare Diagnostics, Newark, DE
Classification
Class II
Status
Terminated
Recall initiated
30 Oct 2013
FDA classified
23 Jan 2014
Reported
29 Jan 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

30 Oct 2013Recall initiated by company
23 Jan 2014Classified by FDA+85 days
29 Jan 2014Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Under extremely rare circumstances, a patient result that had been previously disabled may be released without proper review. (Disabled Test in CentraLink System or Aptio System May Become Enabled in CentraLink System.)

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.

Where it was distributed

US Distribution in the states of: MN, NJ, and NV.

MinnesotaNew JerseyNevada

Questions about this recall

Why was it recalled?

Under extremely rare circumstances, a patient result that had been previously disabled may be released without proper review. (Disabled Test in CentraLink System or Aptio System May Become Enabled in CentraLink System.)

Which lots are affected?

CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 Catalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877, 10815474.

Where was it sold?

US Distribution in the states of: MN, NJ, and NV.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Siemens Healthcare Diagnostics

All 694

Other Medical Software recalls

All 903