CLASS II ONGOING Device recall · Reported 30 Apr 2025 · FDA Enforcement Report

Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount recalled

IHB Operations is recalling Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount, distributed nationwide in the US. The recall was initiated on 27 Mar 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI/DI 10841736116071, All Serial Numbers
Quantity recalled112 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1663-2025
FDA event ID
96581
Recalling firm
IHB Operations, Zwolle, N/A
Classification
Class II
Status
Ongoing
Recall initiated
27 Mar 2025
FDA classified
23 Apr 2025
Reported
30 Apr 2025
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

27 Mar 2025Recall initiated by company
23 Apr 2025Classified by FDA+27 days
30 Apr 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Where it was distributed

US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.

Nationwide AlabamaArizonaCaliforniaFloridaIowaIdahoIllinoisLouisianaMassachusettsMarylandMichiganMinnesotaNorth CarolinaNew HampshireNew JerseyNew MexicoNew YorkPennsylvaniaTennesseeTexas

Questions about this recall

Why was it recalled?

The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.

Which lots are affected?

UDI/DI 10841736116071, All Serial Numbers

Where was it sold?

US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from IHB Operations

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIFreedom Boom recalled by Ihb Operations B.vIHB OperationsDevice Malfunction36 states

Other Patient Monitors recalls

All 1,022