CLASS II Device recall · Reported 29 Jan 2020 · FDA Enforcement Report

Sodium Bicarbonate Mixing and Distribution System recalled by Isopure

Isopure is recalling Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish, distributed in 13 states. The recall was initiated on 19 Apr 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesIsoMix Pro UDI: (01)00852456007085; ISOMIX Serial Number IMP-838703 with Pump SN 180210D18
Quantity recalled1

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0822-2020
FDA event ID
83990
Recalling firm
Isopure, Louisville, KY
Classification
Class II
Status
Terminated
Recall initiated
19 Apr 2019
FDA classified
18 Jan 2020
Reported
29 Jan 2020
Type
Voluntary: Firm initiated
Customers notified by
Telephone
Reason category
Foreign Material

Timeline

19 Apr 2019Recall initiated by company
18 Jan 2020Classified by FDA+274 days
29 Jan 2020Published in enforcement report+11 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3

Where it was distributed

The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.

AlabamaFloridaGeorgiaMassachusettsMarylandMinnesotaNorth CarolinaNew MexicoNew YorkOhioOklahomaTexasVirginia

Questions about this recall

Is Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson recalled?

Yes. Isopure recalled Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson on 19 Apr 2019. The recall is Class II and its status is Terminated. Recall number Z-0822-2020.

Why was it recalled?

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Which lots are affected?

IsoMix Pro UDI: (01)00852456007085; ISOMIX Serial Number IMP-838703 with Pump SN 180210D18

Where was it sold?

The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Isopure

All 5

Other Other Medical Devices recalls

All 13,778