CLASS II Device recall · Reported 30 Oct 2013 · FDA Enforcement Report

Sterrad 100NX Cassettes The Sterrad Sterilization System recalled

Advanced Sterilization Products is recalling Sterrad 100NX Cassettes The Sterrad Sterilization System is a low temperature general, distributed nationwide in the US. The recall was initiated on 18 Sep 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot# 13F050, 13F051, 13F049
Quantity recalled2,995 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0044-2014
FDA event ID
66317
Recalling firm
Advanced Sterilization Products, Irvine, CA
Classification
Class II
Status
Terminated
Recall initiated
18 Sep 2013
FDA classified
22 Oct 2013
Reported
30 Oct 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

18 Sep 2013Recall initiated by company
22 Oct 2013Classified by FDA+34 days
30 Oct 2013Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Where it was distributed

Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.

Which lots are affected?

Lot# 13F050, 13F051, 13F049

Where was it sold?

Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.