CLASS II ONGOING Device recall · Reported 12 Feb 2025 · FDA Enforcement Report

Sysmex PS-10 Sample Preparation System recalled by Sysmex America

Sysmex America is recalling Sysmex PS-10 Sample Preparation System, distributed nationwide in the US. The recall was initiated on 6 Jan 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 04987562501908; Serial Numbers: 1908002275 1908002276 1908009775 1910009588 1910011295 1912002638 2103013913 2108004629 2108004630 2108004631 2108004632 2112005423 2112005425 2203014206 2203014207 2203014208 2205012732 2205012734 2206005973 2206012294 2208015835 2210003722 2210003723 2210003724 2212010048 1903011966 1903011968 1910009589 1912002640 2103013912 2208015834 1903011967
Quantity recalled32 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1113-2025
FDA event ID
96107
Recalling firm
Sysmex America, Lincolnshire, IL
Classification
Class II
Status
Ongoing
Recall initiated
6 Jan 2025
FDA classified
5 Feb 2025
Reported
12 Feb 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

6 Jan 2025Recall initiated by company
5 Feb 2025Classified by FDA+30 days
12 Feb 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.

Where it was distributed

US Nationwide.

Nationwide

Questions about this recall

Why was it recalled?

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

Which lots are affected?

UDI-DI: 04987562501908; Serial Numbers: 1908002275 1908002276 1908009775 1910009588 1910011295 1912002638 2103013913 2108004629 2108004630 2108004631 2108004632 2112005423 2112005425 2203014206 2203014207 2203014208 2205012732 2205012734 2206005973 2206012294 2208015835 2210003722 2210003723 2210003724 2212010048 1903011966 1903011968 1910009589 1912002640 2103013912 2208015834 1903011967

Where was it sold?

US Nationwide.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Sysmex America

All 14

Other Medical Software recalls

All 932