CLASS II Device recall · Reported 8 Nov 2017 · FDA Enforcement Report
Thermocool SF NAV Catheters recalled by Biosense Webster
Biosense Webster is recalling Thermocool SF NAV Catheters. The recall was initiated on 6 Sep 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0058-2018
- FDA event ID
- 78129
- Recalling firm
- Biosense Webster, Irwindale, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 6 Sep 2017
- FDA classified
- 27 Oct 2017
- Reported
- 8 Nov 2017
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Lead & Heavy Metals
- Device type
- Heart & Vascular Devices
Timeline
| 6 Sep 2017 | Recall initiated by company | |
| 27 Oct 2017 | Classified by FDA | +51 days |
| 8 Nov 2017 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
THERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-S
Where it was distributed
US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China
Questions about this recall
Is Thermocool SF NAV Catheters recalled?
Yes. Biosense Webster recalled Thermocool SF NAV Catheters on 6 Sep 2017. The recall is Class II and its status is Terminated. Recall number Z-0058-2018.
Why was it recalled?
Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
Which lots are affected?
17637495L 17638389L 17639706L 17639707L 17639708L 17640039L 17640494L 17640952L 17640040L 17640742L 17640953L 17640041L 17640743L 17640954L 17640490L 17640744L 17643130L 17638118L 17638387L 17638388L
Where was it sold?
US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Biosense Webster
All 32| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Varipulse Bi-Directional Ablation Catheter recalled by Biosense Webster | Biosense Webster | Other | 10 states |
| Device | CLASS II | Carto Vizigo Bi-Directional Guiding Sheath recalled by Biosense Webster | Biosense Webster | Other | Nationwide |
| Device | CLASS II | Carto Vizigo Bi-Directional Guiding Sheath, 5F(Medium) recalled | Biosense Webster | Labeling Errors | Nationwide |
| Device | CLASS II | Carto Vizigo Bi-Directional Guiding Sheath, 5F(Small) recalled | Biosense Webster | Labeling Errors | Nationwide |
| Device | CLASS II | Carto Vizigo 5F Bi-Directional Guiding Sheath recalled | Biosense Webster | Sterility | Nationwide |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |