CLASS II Device recall · Reported 17 Feb 2021 · FDA Enforcement Report

TLAB, Transjugular Liver Biopsy System recalled by Argon Medical Devices

Argon Medical Devices is recalling Argon Medical Devices TLAB, Transjugular Liver Biopsy System, distributed in 46 states. The recall was initiated on 10 Dec 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Number: 1439835,1451429, 1486445
Quantity recalled247 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1075-2021
FDA event ID
87000
Recalling firm
Argon Medical Devices, Athens, TX
Classification
Class II
Status
Terminated
Recall initiated
10 Dec 2020
FDA classified
5 Feb 2021
Reported
17 Feb 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

10 Dec 2020Recall initiated by company
5 Feb 2021Classified by FDA+57 days
17 Feb 2021Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19S, 1/19 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/5f straight catheter, Sterile EO,

Where it was distributed

US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain

AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland

Questions about this recall

Is TLAB, Transjugular Liver Biopsy System recalled?

Yes. Argon Medical Devices recalled TLAB, Transjugular Liver Biopsy System on 10 Dec 2020. The recall is Class II and its status is Terminated. Recall number Z-1075-2021.

Why was it recalled?

Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.

Which lots are affected?

Lot Number: 1439835,1451429, 1486445

Where was it sold?

US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Argon Medical Devices

All 50

Other Heart & Vascular Devices recalls

All 6,199