CLASS II Device recall · Reported 18 Jan 2017 · FDA Enforcement Report
Ultra(R) Sterrad container recalled by Symmetry Surgical
Symmetry Surgical is recalling Ultra(R) Sterrad container. The recall was initiated on 29 Nov 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1011-2017
- FDA event ID
- 75850
- Recalling firm
- Symmetry Surgical, Antioch, TN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 29 Nov 2016
- FDA classified
- 11 Jan 2017
- Reported
- 18 Jan 2017
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Surgical Instruments
Timeline
| 29 Nov 2016 | Recall initiated by company | |
| 11 Jan 2017 | Classified by FDA | +43 days |
| 18 Jan 2017 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Sealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
symmetry surgical ULTRA(R) Sterrad container, REF numbers: C131310-SP, C13134-SP, C13135-SP, C13136-SP, C13139-SP, C13854-SP, C13855-SP, C13856-SP, C181310-SP, C18134-SP, C18135-SP, C18136-SP, C18138-SP, C22854-SP, C22855-SP, C22856-SP, C22858-SP, C24134-SP, C24135-SP, C24136-SP, C24138-SP Product Usage: sterilization container
Where it was distributed
Worldwide Distribution
Questions about this recall
Why was it recalled?
Sealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
Which lots are affected?
All
Where was it sold?
Worldwide Distribution
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
2 other products · FDA event 75850This product is part of a recall event; the main page for the event is Ultra(R) Steam container recalled by Symmetry Surgical.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Ultra(R) Steam container recalled by Symmetry Surgical | Symmetry Surgical | Sterility | TN | |
| CLASS II | Ultra(R) Gravity container recalled by Symmetry Surgical | Symmetry Surgical | Sterility | TN |
More recalls from Symmetry Surgical
All 4| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Ultra(R) Gravity container recalled by Symmetry Surgical | Symmetry Surgical | Sterility | TN |
| Device | CLASS II | Ultra(R) Steam container recalled by Symmetry Surgical | Symmetry Surgical | Sterility | TN |
| Device | CLASS II | Disposable Towel Clamp recalled by Symmetry Surgical | Symmetry Surgical | Sterility | 18 states |
Other Surgical Instruments recalls
All 3,809| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: IV Start KIT recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: Open Heart Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: GYN Robotic CDS, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Linvatec Genesys CrossFT Suture Anchor with Disposable Driver recalled | ConMed | Other | Nationwide |