CLASS II ONGOING Device recall · Reported 25 Jun 2025 · FDA Enforcement Report

Uroskop Omnia recalled by Siemens Medical Solutions USA

Siemens Medical Solutions USA is recalling Uroskop Omnia, distributed nationwide in the US. The recall was initiated on 4 Jun 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number: 10094910. No UDI. Serial Numbers: 3054, 3029, 3062, 3169, 3402, 3103, 3053, 3055, 3072, 3019, 3048, 3403, 3058, 1045, 2014, 3091, 3013, 3023, 3115, 3051, 3417, 3160, 2033, 3163, 3052, 3166, 3099, 3138, 3157, 3038, 3131, 3037, 3117, 3118, 3096, 3123, 3124, 2022, 2032, 2031, 3134, 3106, 3415, 3042, 3014, 3012, 3108, 3105, 3057, 2008, 3122, 3034, 3151, 3414, 2030, 3021, 3119, 3086, 3073, 3089, 3036, 3031, 3409, 3064, 3153, 3095, 3145, 3079, 3017, 3025, 3076, 2046, 2047, 2055, 2054, 2049, 3416, 2015, 1041, 3155, 3418, 3081, 3084, 2045, 3015, 3125, 3132, 3143, 3085, 3405, 2041, 2009, 3410, 2029, 3404.
Quantity recalled95 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1981-2025
FDA event ID
97050
Recalling firm
Siemens Medical Solutions USA, Malvern, PA
Classification
Class II
Status
Ongoing
Recall initiated
4 Jun 2025
FDA classified
17 Jun 2025
Reported
25 Jun 2025
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Foreign Material

Timeline

4 Jun 2025Recall initiated by company
17 Jun 2025Classified by FDA+13 days
25 Jun 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

UROSKOP Omnia. Model Number: 10094910

Where it was distributed

Nationwide distribution.

Nationwide

Questions about this recall

Is Uroskop Omnia recalled?

Yes. Siemens Medical Solutions USA recalled Uroskop Omnia on 4 Jun 2025. The recall is Class II and its status is Ongoing. Recall number Z-1981-2025.

Why was it recalled?

A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.

Which lots are affected?

Model Number: 10094910. No UDI. Serial Numbers: 3054, 3029, 3062, 3169, 3402, 3103, 3053, 3055, 3072, 3019, 3048, 3403, 3058, 1045, 2014, 3091, 3013, 3023, 3115, 3051, 3417, 3160, 2033, 3163, 3052, 3166, 3099, 3138, 3157, 3038, 3131, 3037, 3117, 3118, 3096, 3123, 3124, 2022, 2032, 2031, 3134, 3106, 3415, 3042, 3014, 3012, 3108, 3105, 3057, 2008, 3122, 3034, 3151, 3414, 2030, 3021, 3119, 3086, 3073, 3089, 3036, 3031, 3409, 3064, 3153, 3095, 3145, 3079, 3017, 3025, 3076, 2046, 2047, 2055, 2054, 2049, 3416, 2015, 1041, 3155, 3418, 3081, 3084, 2045, 3015, 3125, 3132, 3143, 3085, 3405, 2041, 2009, 3410, 2029, 3404.

Where was it sold?

Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Siemens Medical Solutions USA

All 796

Other Other Medical Devices recalls

All 13,778