CLASS II ONGOING Device recall · Reported 17 Jan 2024 · FDA Enforcement Report

Vader pedicle system, torque wrench recalled by Icotec Ag

Icotec is recalling Vader pedicle system, torque wrench, distributed in 10 states. The recall was initiated on 28 Nov 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes19/01-001, 19/01-013, 19/01-014, 19/01-015, 190830-10, 190830-23, 190830-42, 190830-44, 190830-45, 190830-50, 190830-51, 190830-52, 200131-54, 200131-56, 200131-62, 200131-64, 200131-68, 200131-69, 200131-70, 200131-72, 200131-73, 200131-74, 200221-040R1, 200424-01, 200424-03, 200424-04, 200424-05, 200424-06, 200424-10, 200424-11, 200424-12, 200424-13, 200424-14, 200424-15, 200424-16, 200424-17, 200424-20, 200424-21, 200424-22, 200424-24, 200424-25, 200424-27, 200424-29, 200424-32, 200424-34, 200424-35, 200605-051R1, 200605-052R1, 201002-073R1, 210915-005, 210915-009, 210915-010, 210915-011, 210915-012, 220610-001, 220610-003, 220610-008, 220610-019, 220610-020, 220610-021, 220610-025, 220610-026, 220610-030, 230113-003, 230113-005, 230113-007, 230113-008, 230113-009, 230113-010, 230113-011, 230113-012, 230113-013, 230113-014, 230224-029, 230224-034, 230426-038, 230426-039, 230426-042, 230426-043
Quantity recalled79 devices (US only)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0700-2024
FDA event ID
93577
Recalling firm
Icotec, Altstatten Sg
Classification
Class II
Status
Ongoing
Recall initiated
28 Nov 2023
FDA classified
10 Jan 2024
Reported
17 Jan 2024
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction
Device type
Spinal Implants

Timeline

28 Nov 2023Recall initiated by company
10 Jan 2024Classified by FDA+43 days
17 Jan 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VADER pedicle system, torque wrench, Catalog Number 42-702

Where it was distributed

US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA

ArizonaCaliforniaColoradoDelawareMichiganNorth CarolinaNew YorkOhioPennsylvaniaTexas

Questions about this recall

Why was it recalled?

icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.

Which lots are affected?

19/01-001, 19/01-013, 19/01-014, 19/01-015, 190830-10, 190830-23, 190830-42, 190830-44, 190830-45, 190830-50, 190830-51, 190830-52, 200131-54, 200131-56, 200131-62, 200131-64, 200131-68, 200131-69, 200131-70, 200131-72, 200131-73, 200131-74, 200221-040R1, 200424-01, 200424-03, 200424-04, 200424-05, 200424-06, 200424-10, 200424-11, 200424-12, 200424-13, 200424-14, 200424-15, 200424-16, 200424-17, 200424-20, 200424-21, 200424-22, 200424-24, 200424-25, 200424-27, 200424-29, 200424-32, 200424-34, 200424-35, 200605-051R1, 200605-052R1, 201002-073R1, 210915-005, 210915-009, 210915-010, 210915-011, 210915-012, 220610-001, 220610-003, 220610-008, 220610-019, 220610-020, 220610-021, 220610-025, 220610-026, 220610-030, 230113-003, 230113-005, 230113-007, 230113-008, 230113-009, 230113-010, 230113-011, 230113-012, 230113-013, 230113-014, 230224-029, 230224-034, 230426-038, 230426-039, 230426-042, 230426-043

Where was it sold?

US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 93577
ReportedClassRecallCompanyReasonWhere
CLASS IIVader pedicle system, torque wrench recalled by Icotec AgIcotecDevice Malfunction10 states

More recalls from Icotec

All 3

Other Spinal Implants recalls

All 1,157