CLASS I ONGOING Device recall · Reported 19 Jan 2022 · FDA Enforcement Report
Vaporizer Sevoflurane Maquet Filling recalled by Getinge Usa Sales
Getinge Usa Sales is recalling Vaporizer Sevoflurane Maquet Filling, distributed nationwide in the US. The recall was initiated on 9 Dec 2021 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0468-2022
- FDA event ID
- 89125
- Recalling firm
- Getinge Usa Sales, Wayne, NJ
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 9 Dec 2021
- FDA classified
- 10 Jan 2022
- Reported
- 19 Jan 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Foreign Material
- Device type
- Other Medical Devices
Timeline
| 9 Dec 2021 | Recall initiated by company | |
| 10 Jan 2022 | Classified by FDA | +32 days |
| 19 Jan 2022 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
Vaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia System 6887700 07325710009765 Flow-e Anesthesia System 6887900 07325710010457 Flow-i Anesthesia System C20 6888520 07325710010617 Flow-i Anesthesia System C30 6888530 07325710010624 Flow-i Anesthesia System C40 6888540 07325710010631 Flow-i C20 Anesthesia System 6677200 07325710001349 Flow-i C30 Anesthesia System 6677300 07325710001349 Flow-i C40 Anesthesia System 6677400 07325710001318
Where it was distributed
Worldwide distribution - US Nationwide distribution in the states of VA and MO. The countries of Austria, Bahrain, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Estonia, Iraq, Israel, Italy, Japan, Kenya, Malaysia, Mexico, Morocco, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Questions about this recall
Why was it recalled?
A potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.
Which lots are affected?
UDI: 07325710008430 Serial Nos. 15580 15890 15924 16284 16637 16837 16970 15582 15891 15925 16285 16638 16840 16971 15721 15894 16268 16345 16639 16842 15722 15914 16269 16352 16640 16843 15725 15917 16270 16354 16690 16844 15726 15919 16271 16632 16828 16961 15728 15920 16272 16633 16832 16964 15729 15921 16273 16634 16833 16965 15731 15922 16276 16635 16834 16966 15732 15923 16283 16636 16836 16968
Where was it sold?
Worldwide distribution - US Nationwide distribution in the states of VA and MO. The countries of Austria, Bahrain, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Estonia, Iraq, Israel, Italy, Japan, Kenya, Malaysia, Mexico, Morocco, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Getinge Usa Sales
All 47| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Vaporizer Sevoflurane, Quik-Fil recalled by Getinge Usa Sales | Getinge Usa Sales | Other | CT, KY |
| Device | CLASS II | Meera CL mobile operating table - Designed recalled by Getinge Usa Sales | Getinge Usa Sales | Other | Nationwide |
| Device | CLASS II | Meera CL mobile operating table - Designed recalled by Getinge Usa Sales | Getinge Usa Sales | Other | Nationwide |
| Device | CLASS II | Meera CL mobile operating table - Designed recalled by Getinge Usa Sales | Getinge Usa Sales | Other | Nationwide |
| Device | CLASS II | Meera CL mobile operating table - Designed recalled by Getinge Usa Sales | Getinge Usa Sales | Other | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |