CLASS II Device recall · Reported 27 Jun 2018 · FDA Enforcement Report

VNS Therapy Programmer recalled by Cyberonics

Cyberonics is recalling VNS Therapy Programmer, distributed in 21 states. The recall was initiated on 13 Nov 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: FMT77F2, FKR57F2, FCR77F2, CJR77F2, BNT77F2, BNS77F2, 7XR77F2, 7XP57F2, 79R77F2, 90S77F2, 97Q57F2, 6NT77F2, 9GP57F2, 6KQ57F2, B7Q57F2, 57P57F2, 55T77F2, J6R57F2, HWR77F2, 1G7G1F2, 1JR57F2, 9XP57F2, 4LR77F2, 4GT77F2, 4GR57F2, 2DT77F2, G2S77F2, FVQ77F2, G5T77F2, FDQ57F2, F5R57F2, F2R57F2, F1Q57F2, GHS77F2, DVR77F2, DNP57F2, DMYB7F2, DJS77F2, D1R57F2, CVQ57F2, H4R57F2, 85S77F2, 89T77F2, 8LYB7F2, 97YB7F2, 6MQ57F2, 6CQ57F2, 5HS77F2, JLQ57F2, JGS77F2, 11R57F2, 1TQ57F2, 1YR77F2, 2DR77F2, ,FXP57F2 , FTR77F2, FPQ57F2, FNQ57F2, FLYB7F2, FFP57F2, GGT77F2, GSQ57F2, CMS77F2, H1R77F2, H3R57F2, 7LS77F2, 9KR57F2, 9NS77F2, 65R57F2, B8R77F2, B9T77F2, 5D0C7F2, B9YB7F2, BJR57F2, JCQ57F2, J9YB7F2, HMYB7F2, CDQ57F2, FXQ57F2, G5S77F2, G3R77F2, FRQ57F2, G7YB7F2, FLS77F2, G90C7F2, FGR77F2, FFT77F2, G9Q57F2, FCT77F2, G9R57F2, F9YB7F2, F9S77F2, F8T77F2, GDQ57F2, F7R57F2, F4T77F2, DWR77F2, DTR77F2, DSQ57F2, GMQ57F2, DLS77F2, DH0C7F2, DGS77F2, DGP57F2, GZR77F2, D1Q57F2, CSQ57F2, CLS77F2, C9YB7F2, C70C7F2, H1R57F2, C1R77F2, 7MS77F2, 78YB7F2, 9SQ57F2, 5VQ57F2, 5RQ57F2, B60C7F2, 5NQ57F2, 5MR77F2, 5KS77F2, B8T77F2, BHP57F2, BJR77F2, 4JS77F2, H4T77F2, JZQ57F2, JMS77F2, JMQ57F2, JG0C7F2, JBQ57F2, J8T77F2, ,HZQ57F2, J7R57F2, HVR77F2, HTP57F2, HNS77F2, HMQ57F2, HKQ57F2, H7S77F2, D3R57F2, 1CQ57F2, 1HT77F2, 1MYB7F2, 2CP57F2, 2LYB7F2, CLYB7F2, CMYB7F2, B4R77F2, B2S77F2, B0R57F2, 9QS77F2, D8Q57F2, 960C7F2, 95T77F2, 8JR77F2, 8HR57F2, BXP57F2, BWR77F2, BTQ77F2, C8YB7F2, C60C7F2, BLR77F2, BLQ57F2, BGR77F2, BGP57F2, CHQ57F2, CKR57F2, CPP57F2, CSR77F2, CTQ77F2, CXR77F2, 9MS77F2, 9JR77F2, 9DQ57F2, D7S77F2, 94T77F2, 8PQ57F2, BWQ77F2, BSR77F2, BNQ57F2, C9R57F2, C8R57F2, C5R57F2, CF0C7F2, C3R57F2, CFP57F2, BDT77F2, B9S77F2, B5R57F2, CPQ57F2, B4R57F2, D2R57F2, 9KQ57F2, D5R77F2, 9F0C7F2, 980C7F2, D9YB7F2, 8WR77F2, DB0C7F2, 9P7G1F2, D3S77F2, D6R57F2, D7R57F2, 9CR77F2, 9G0C7F2, 96R57F2, DFP57F2, C0S77F2, BN7G1F2, BQ7G1F2, C8S77F2, CB0C7F2, CCR77F2, BHR57F2, CHP57F2, BCR77F2, BBR77F2, BB0C7F2, B90C7F2, B70C7F2, B2R57F2, 9VP57F2, 9TQ57F2, 9GR57F2, 9FP57F2, 9BS77F2, ,98R57F2 ,D80C7F2, 970C7F2, D8YB7F2, 95R77F2, 8YR77F2, C1R57F2, C80C7F2, C3R77F2, CNS77F2, 9VQ77F2, D4T77F2, 9DR77F2, 91R57F2, 8VQ77F2, 8RQ57F2, DBS77F2, 8JS77F2, C0R57F2, BTR77F2, BMYB7F2, BKR77F2, BBP57F2 and B8S77F2
Quantity recalled252

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2255-2018
FDA event ID
80062
Recalling firm
Cyberonics, Houston, TX
Classification
Class II
Status
Terminated
Recall initiated
13 Nov 2017
FDA classified
19 Jun 2018
Reported
27 Jun 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Lead & Heavy Metals

Timeline

13 Nov 2017Recall initiated by company
19 Jun 2018Classified by FDA+218 days
27 Jun 2018Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).

Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.

Product as listed by the FDA

VNS Therapy Programmer, Model 3000, v1.0 System

Where it was distributed

AL, CA, CO, DE, FL, GA, ID, IL, IN, ME, MO, MS, NC, NJ, NY, PA, TN, TN, TX, UT, WA and WI

AlabamaCaliforniaColoradoDelawareFloridaGeorgiaIdahoIllinoisIndianaMaineMissouriMississippiNorth CarolinaNew JerseyNew YorkPennsylvaniaTennesseeTexasUtahWashington

Questions about this recall

Is VNS Therapy Programmer recalled?

Yes. Cyberonics recalled VNS Therapy Programmer on 13 Nov 2017. The recall is Class II and its status is Terminated. Recall number Z-2255-2018.

Why was it recalled?

Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).

Which lots are affected?

Serial Numbers: FMT77F2, FKR57F2, FCR77F2, CJR77F2, BNT77F2, BNS77F2, 7XR77F2, 7XP57F2, 79R77F2, 90S77F2, 97Q57F2, 6NT77F2, 9GP57F2, 6KQ57F2, B7Q57F2, 57P57F2, 55T77F2, J6R57F2, HWR77F2, 1G7G1F2, 1JR57F2, 9XP57F2, 4LR77F2, 4GT77F2, 4GR57F2, 2DT77F2, G2S77F2, FVQ77F2, G5T77F2, FDQ57F2, F5R57F2, F2R57F2, F1Q57F2, GHS77F2, DVR77F2, DNP57F2, DMYB7F2, DJS77F2, D1R57F2, CVQ57F2, H4R57F2, 85S77F2, 89T77F2, 8LYB7F2, 97YB7F2, 6MQ57F2, 6CQ57F2, 5HS77F2, JLQ57F2, JGS77F2, 11R57F2, 1TQ57F2, 1YR77F2, 2DR77F2, ,FXP57F2 , FTR77F2, FPQ57F2, FNQ57F2, FLYB7F2, FFP57F2, GGT77F2, GSQ57F2, CMS77F2, H1R77F2, H3R57F2, 7LS77F2, 9KR57F2, 9NS77F2, 65R57F2, B8R77F2, B9T77F2, 5D0C7F2, B9YB7F2, BJR57F2, JCQ57F2, J9YB7F2, HMYB7F2, CDQ57F2, FXQ57F2, G5S77F2, G3R77F2, FRQ57F2, G7YB7F2, FLS77F2, G90C7F2, FGR77F2, FFT77F2, G9Q57F2, FCT77F2, G9R57F2, F9YB7F2, F9S77F2, F8T77F2, GDQ57F2, F7R57F2, F4T77F2, DWR77F2, DTR77F2, DSQ57F2, GMQ57F2, DLS77F2, DH0C7F2, DGS77F2, DGP57F2, GZR77F2, D1Q57F2, CSQ57F2, CLS77F2, C9YB7F2, C70C7F2, H1R57F2, C1R77F2, 7MS77F2, 78YB7F2, 9SQ57F2, 5VQ57F2, 5RQ57F2, B60C7F2, 5NQ57F2, 5MR77F2, 5KS77F2, B8T77F2, BHP57F2, BJR77F2, 4JS77F2, H4T77F2, JZQ57F2, JMS77F2, JMQ57F2, JG0C7F2, JBQ57F2, J8T77F2, ,HZQ57F2, J7R57F2, HVR77F2, HTP57F2, HNS77F2, HMQ57F2, HKQ57F2, H7S77F2, D3R57F2, 1CQ57F2, 1HT77F2, 1MYB7F2, 2CP57F2, 2LYB7F2, CLYB7F2, CMYB7F2, B4R77F2, B2S77F2, B0R57F2, 9QS77F2, D8Q57F2, 960C7F2, 95T77F2, 8JR77F2, 8HR57F2, BXP57F2, BWR77F2, BTQ77F2, C8YB7F2, C60C7F2, BLR77F2, BLQ57F2, BGR77F2, BGP57F2, CHQ57F2, CKR57F2, CPP57F2, CSR77F2, CTQ77F2, CXR77F2, 9MS77F2, 9JR77F2, 9DQ57F2, D7S77F2, 94T77F2, 8PQ57F2, BWQ77F2, BSR77F2, BNQ57F2, C9R57F2, C8R57F2, C5R57F2, CF0C7F2, C3R57F2, CFP57F2, BDT77F2, B9S77F2, B5R57F2, CPQ57F2, B4R57F2, D2R57F2, 9KQ57F2, D5R77F2, 9F0C7F2, 980C7F2, D9YB7F2, 8WR77F2, DB0C7F2, 9P7G1F2, D3S77F2, D6R57F2, D7R57F2, 9CR77F2, 9G0C7F2, 96R57F2, DFP57F2, C0S77F2, BN7G1F2, BQ7G1F2, C8S77F2, CB0C7F2, CCR77F2, BHR57F2, CHP57F2, BCR77F2, BBR77F2, BB0C7F2, B90C7F2, B70C7F2, B2R57F2, 9VP57F2, 9TQ57F2, 9GR57F2, 9FP57F2, 9BS77F2, ,98R57F2 ,D80C7F2, 970C7F2, D8YB7F2, 95R77F2, 8YR77F2, C1R57F2, C80C7F2, C3R77F2, CNS77F2, 9VQ77F2, D4T77F2, 9DR77F2, 91R57F2, 8VQ77F2, 8RQ57F2, DBS77F2, 8JS77F2, C0R57F2, BTR77F2, BMYB7F2, BKR77F2, BBP57F2 and B8S77F2

Where was it sold?

AL, CA, CO, DE, FL, GA, ID, IL, IN, ME, MO, MS, NC, NJ, NY, PA, TN, TN, TX, UT, WA and WI

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.