CLASS I ONGOING Device recall · Reported 28 Oct 2026 · FDA Enforcement Report

Vocsn (Japan) recalled by Ventec Life Systems

Ventec Life Systems is recalling Vocsn (Japan), distributed nationwide in the US. The recall was initiated on 17 Sep 2026 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesAll Serial Numbers: UDI: 00850018761130
Quantity recalled100 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0127-2027
FDA event ID
99835
Recalling firm
Ventec Life Systems, Bothell, WA
Classification
Class I
Status
Ongoing
Recall initiated
17 Sep 2026
FDA classified
20 Oct 2026
Reported
28 Oct 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

17 Sep 2026Recall initiated by company
20 Oct 2026Classified by FDA+33 days
28 Oct 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.

Where it was distributed

Distribution US nationwide including Puerto Rico.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

Which lots are affected?

All Serial Numbers: UDI: 00850018761130

Where was it sold?

Distribution US nationwide including Puerto Rico.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 3 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

11 other products · FDA event 99835

More recalls from Ventec Life Systems

All 20

Other CPAP & Ventilators recalls