CLASS II ONGOING Device recall · Reported 11 Feb 2026 · FDA Enforcement Report
Volk Single Use Vitrectomy Lenses Flat Vitrectomy Lens recalled
Vortex Surgical is recalling Volk Single Use Vitrectomy Lenses Flat Vitrectomy Lens, distributed nationwide in the US. The recall was initiated on 16 Dec 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1244-2026
- FDA event ID
- 98255
- Recalling firm
- Vortex Surgical, Saint Charles, MO
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 16 Dec 2025
- FDA classified
- 4 Feb 2026
- Reported
- 11 Feb 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Other Medical Devices
Timeline
| 16 Dec 2025 | Recall initiated by company | |
| 4 Feb 2026 | Classified by FDA | +50 days |
| 11 Feb 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD
Where it was distributed
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
Nationwide AlabamaArkansasArizonaCaliforniaConnecticutFloridaHawaiiIowaIllinoisIndianaKentuckyMassachusettsMichiganMinnesotaMississippiNorth CarolinaNebraskaNew JerseyNew MexicoNew York
Questions about this recall
Why was it recalled?
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
Which lots are affected?
1. Flat Vitrectomy Lens Pouch UDI 810123482290 Box UDI 810123482306 Lots 2410036 2411020 2. Volk Magnifying Lens Pouch UDI 810123482825 Box UDI 810123482832 Lot 2411019
Where was it sold?
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
12 other products · FDA event 98255This product is part of a recall event; the main page for the event is Convenience Kits recalled by Vortex Surgical.
More recalls from Vortex Surgical
All 15| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Lindsell Sutured IOL Marker, VS0390 - Scleral Marker recalled | Vortex Surgical | Sterility | Nationwide |
| Device | CLASS II | Rumex Disposable Diamond Dusted ILM Elevator recalled by Vortex Surgical | Vortex Surgical | Sterility | Nationwide |
| Device | CLASS II | Vortex Disposable Forceps and Cannula recalled by Vortex Surgical | Vortex Surgical | Sterility | Nationwide |
| Device | CLASS II | Subretinal Injection Cannula VS0220.25 recalled by Vortex Surgical | Vortex Surgical | Sterility | Nationwide |
| Device | CLASS II | Tecfen Retractable Membrane Polisher, QTPR1267 recalled | Vortex Surgical | Sterility | Nationwide |
Other Other Medical Devices recalls
All 13,969| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Philips Azurion 3 M15 recalled by Philips Medical Systems Nederl | Philips Medical Systems Nederl | Other | Nationwide |
| Device | CLASS II | Orthoscan Versa Accessory Cart recalled by Ziehm-orthoscan | Ziehm-orthoscan | Device Malfunction | Nationwide |
| Device | CLASS II | Philips Azurion 7 B12 recalled by Philips Medical Systems Nederl | Philips Medical Systems Nederl | Other | Nationwide |
| Device | CLASS II | Philips Azurion 7 M20 recalled by Philips Medical Systems Nederl | Philips Medical Systems Nederl | Other | Nationwide |