CLASS II ONGOING Device recall · Reported 28 Dec 2022 · FDA Enforcement Report

2008T and 2008TBluestar Hemodialysis Machines recalled

Fresenius Medical Care Holdings is recalling 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic, distributed nationwide in the US. The recall was initiated on 27 Jul 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware issue in SW versions 2.72 and earlier with CDX. UDI: (1) 00840861100897; (2) 00840861100910; (3) 00840861102082; (4) 00840861102099
Quantity recalled2008T with CDX= 70,552 2008T Bluestar with CDX= 25,205

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0734-2023
FDA event ID
91154
Recalling firm
Fresenius Medical Care Holdings, Waltham, MA
Classification
Class II
Status
Ongoing
Recall initiated
27 Jul 2022
FDA classified
20 Dec 2022
Reported
28 Dec 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

27 Jul 2022Recall initiated by company
20 Dec 2022Classified by FDA+146 days
28 Dec 2022Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

Which lots are affected?

Software issue in SW versions 2.72 and earlier with CDX. UDI: (1) 00840861100897; (2) 00840861100910; (3) 00840861102082; (4) 00840861102099

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Fresenius Medical Care Holdings

All 165

Other Other Medical Devices recalls

All 13,778