CLASS II Device recall · Reported 7 Mar 2018 · FDA Enforcement Report
2008T, Hemodialysis Delivery System Product Indicated recalled
Fresenius Medical Care Renal Therapies Group is recalling 2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis, distributed nationwide in the US. The recall was initiated on 5 Oct 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0829-2018
- FDA event ID
- 79239
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, Waltham, MA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 5 Oct 2017
- FDA classified
- 1 Mar 2018
- Reported
- 7 Mar 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Telephone
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 5 Oct 2017 | Recall initiated by company | |
| 1 Mar 2018 | Classified by FDA | +147 days |
| 7 Mar 2018 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.
Where it was distributed
US Nationwide Distribution
Questions about this recall
Why was it recalled?
While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
Which lots are affected?
2008T machines: 190766 2008T Hemodialysis System w/Bibag 190895 2008T GEN 2 Bibag without CDX 190766 = 00840861100910 (+Serial Number) 190895 = 00840861100927 (+Serial Number)ux=&Su
Where was it sold?
US Nationwide Distribution
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Fresenius Medical Care Renal Therapies Group
All 40| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Drug | CLASS II | Delflex Peritoneal Dialysis Solution with 2.5% Dextrose recalledDelflex | Fresenius Medical Care Renal Therapies Group | Sterility | Nationwide |
| Drug | CLASS II | Delflex Peritoneal Dialysis Solution with 1.5% Dextrose recalledDelflex | Fresenius Medical Care Renal Therapies Group | Sterility | Nationwide |
| Drug | CLASS II | Delflex Peritoneal Dialysis Solution with 1.5% Dextrose recalledDelflex | Fresenius Medical Care Renal Therapies Group | Sterility | Nationwide |
| Drug | CLASS II | Delflex Peritoneal Dialysis Solution with 2.5% Dextrose recalledDelflex | Fresenius Medical Care Renal Therapies Group | Sterility | Nationwide |
| Drug | CLASS II | Delflex Peritoneal Dialysis Solution with 4.25% Dextrose recalledDelflex | Fresenius Medical Care Renal Therapies Group | Sterility | Nationwide |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |