CLASS II Device recall · Reported 7 Mar 2018 · FDA Enforcement Report

2008T, Hemodialysis Delivery System Product Indicated recalled

Fresenius Medical Care Renal Therapies Group is recalling 2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis, distributed nationwide in the US. The recall was initiated on 5 Oct 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes2008T machines: 190766 2008T Hemodialysis System w/Bibag 190895 2008T GEN 2 Bibag without CDX 190766 = 00840861100910 (+Serial Number) 190895 = 00840861100927 (+Serial Number)ux=&Su
Quantity recalled15 machines

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0829-2018
FDA event ID
79239
Classification
Class II
Status
Terminated
Recall initiated
5 Oct 2017
FDA classified
1 Mar 2018
Reported
7 Mar 2018
Type
Voluntary: Firm initiated
Customers notified by
Telephone
Reason category
Device Malfunction

Timeline

5 Oct 2017Recall initiated by company
1 Mar 2018Classified by FDA+147 days
7 Mar 2018Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.

Which lots are affected?

2008T machines: 190766 2008T Hemodialysis System w/Bibag 190895 2008T GEN 2 Bibag without CDX 190766 = 00840861100910 (+Serial Number) 190895 = 00840861100927 (+Serial Number)ux=&Su

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.