CLASS III ONGOING Device recall · Reported 14 Jul 2021 · FDA Enforcement Report

ACE Control Set - IVD Control set recalled by Sentinel CH SpA

Sentinel CH SpA is recalling ACE Control Set - IVD Control set for the quantitative determination of Angiotensin, distributed nationwide in the US. The recall was initiated on 30 Apr 2021 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesLot Numbers/ Exp. Date: 80516 31/07/2021; 80635 30/09/2021; 90092 31/12/2021; 90599 31/12/2021; 00118 31/12/2021; 00360 31/12/2021; 00610 31/12/2021; 00702 31/12/2021. UDI: 01)08058056681775( 17)210731( 10)80516(240) 1667001 {01)08058056681775( 17)210930(10)80635{240) 1667001 {01)08058056681775( 17)211231(10)90092(240) 1667001 {01)08058056681775( 17)211231(10)90599{240) 1667001 {01)08058056681775( 17)211231( 10)00118(240) 1667001 {01)08058056681775( 17)211231(10)00360(240) 1667001 {01)08058056681775( 17)211231( 10)00610(240) 1667001 {01)08058056681775( 17)211231(10)00702{240) 1667001
Quantity recalled1898 kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2012-2021
FDA event ID
88040
Recalling firm
Sentinel CH SpA, Milan
Classification
Class III
Status
Ongoing
Recall initiated
30 Apr 2021
FDA classified
6 Jul 2021
Reported
14 Jul 2021
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Potency & Assay

Timeline

30 Apr 2021Recall initiated by company
6 Jul 2021Classified by FDA+67 days
14 Jul 2021Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Claim for stability after reconstitution from lyophilized to liquid form does not meet the requirements. New IFU instructions for handling the product provided to end users.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

ACE Control Set - IVD Control set for the quantitative determination of Angiotensin Converting Enzyme (ACE) REF 1667001

Where it was distributed

Worldwide distribution - US Nationwide and the countries of AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, CANADA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ISRAEL, ITALY, JORDAN, NEW ZEALAND, POLAND, PORTUGAL, SERBIA, SOUTH AFRICA, SPAIN, SWEDEN, THE NETHERLANDS, TURKEY, U.A.E., UK.

Nationwide

Questions about this recall

Is ACE Control Set - IVD Control set for the quantitative determination of Angiotensin recalled?

Yes. Sentinel CH SpA recalled ACE Control Set - IVD Control set for the quantitative determination of Angiotensin on 30 Apr 2021. The recall is Class III and its status is Ongoing. Recall number Z-2012-2021.

Why was it recalled?

Claim for stability after reconstitution from lyophilized to liquid form does not meet the requirements. New IFU instructions for handling the product provided to end users.

Which lots are affected?

Lot Numbers/ Exp. Date: 80516 31/07/2021; 80635 30/09/2021; 90092 31/12/2021; 90599 31/12/2021; 00118 31/12/2021; 00360 31/12/2021; 00610 31/12/2021; 00702 31/12/2021. UDI: 01)08058056681775( 17)210731( 10)80516(240) 1667001 {01)08058056681775( 17)210930(10)80635{240) 1667001 {01)08058056681775( 17)211231(10)90092(240) 1667001 {01)08058056681775( 17)211231(10)90599{240) 1667001 {01)08058056681775( 17)211231( 10)00118(240) 1667001 {01)08058056681775( 17)211231(10)00360(240) 1667001 {01)08058056681775( 17)211231( 10)00610(240) 1667001 {01)08058056681775( 17)211231(10)00702{240) 1667001

Where was it sold?

Worldwide distribution - US Nationwide and the countries of AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, CANADA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ISRAEL, ITALY, JORDAN, NEW ZEALAND, POLAND, PORTUGAL, SERBIA, SOUTH AFRICA, SPAIN, SWEDEN, THE NETHERLANDS, TURKEY, U.A.E., UK.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.