CLASS II Device recall · Reported 11 Sep 2013 · FDA Enforcement Report
Advia Chemistry A1c_3 Calibrators recalled
Siemens Healthcare Diagnostics is recalling Advia Chemistry A1c_3 Calibrators, distributed nationwide in the US. The recall was initiated on 30 Jul 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2135-2013
- FDA event ID
- 65868
- Recalling firm
- Siemens Healthcare Diagnostics, Tarrytown, NY
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 30 Jul 2013
- FDA classified
- 3 Sep 2013
- Reported
- 11 Sep 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Hip & Knee Implants
Timeline
| 30 Jul 2013 | Recall initiated by company | |
| 3 Sep 2013 | Classified by FDA | +35 days |
| 11 Sep 2013 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Siemens confirmed that the ADVIA¿ Chemistry Systems HbA1c method exhibits a positive bias of up to 12% for patient and CAP survey samples due to over-recovery when using A1c_3 Calibrator lots 1MD014, 1BD063, and 2GD014. The established ranges of CAP Survey Samples GH2-04, GH2-05 and GH2-06 and IFCC sample recoveries use ¿ 7% from the established mean as their acceptance criteria. This is the criterion that was used during testing to confirm the customer complaints. Depending upon quality control limits this issue may not have been detected..
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
ADVIA Chemistry A1c_3 Calibrators, Catalog Number 10491408, IVD --- COMMON/USUAL NAME: ADVIA Chemistry A1c_3 Calibrators Shipping or unit package: Level 1: 2 x 0.5 g; Level 2: 2 x 0.5 g; Level 3: 2 x 0.5 g; Level 4: 2 x 0.5 g -- Siemens Healthcare Diagnostics, Inc. Tarrytown, NY 10591 --- System Information: The ADVIA Chemistry systems are automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine in random access and batch modes for in vitro diagnostic use. For in vitro diagnostic use in the calibration of A1c_3 (A1c_3 and A1c_3M) and Total Hemoglobin_3 (tHb_3 and tHb_3M) methods on ADVIA¿ Chemistry systems.
Where it was distributed
Worldwide Distribution - USA (nationwide) Puerto Rico and internationally to Argentina, Austria, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hungary, Indonesia, Italy, Kazakhstan, Lithuania, Norway, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, South Africa, South Korea, Thailand, Turkey, Uruguay, Vatican City, and Vietnam.
Questions about this recall
Why was it recalled?
Siemens confirmed that the ADVIA¿ Chemistry Systems HbA1c method exhibits a positive bias of up to 12% for patient and CAP survey samples due to over-recovery when using A1c_3 Calibrator lots 1MD014, 1BD063, and 2GD014. The established ranges of CAP Survey Samples GH2-04, GH2-05 and GH2-06 and IFCC sample recoveries use ¿ 7% from the established mean as their acceptance criteria. This is the criterion that was used during testing to confirm the customer complaints. Depending upon quality control limits this issue may not have been detected..
Which lots are affected?
Lots 1MD014 (exp. date 5/01/2013), 1BD063 (exp. 5/01/2013), 2GD014 (exp. 10/01/2014)
Where was it sold?
Worldwide Distribution - USA (nationwide) Puerto Rico and internationally to Argentina, Austria, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Guatemala, Hungary, Indonesia, Italy, Kazakhstan, Lithuania, Norway, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, South Africa, South Korea, Thailand, Turkey, Uruguay, Vatican City, and Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Siemens Healthcare Diagnostics
All 694| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS III | Syva Emit Disopyramide Assay Enzyme Immunoassay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Atellica CH Urine Albumin recalled by Siemens Healthcare Diagnostics | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS III | Syva Emit tox Serum Tricyclic Antidepressant recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
Other Hip & Knee Implants recalls
All 3,097| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline convenience kits labeled as: Femoral Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Arvis Surgical Navigation System Gen recalled by Kico Knee Innovation | Kico Knee Innovation | Device Malfunction | 19 states |
| Device | CLASS II | Veritas Standard Reverse Total Shoulder Veritas Standard recalled | restor3d | Device Malfunction | Nationwide |
| Device | CLASS II | Endo-Model, Replacement Set, Tibial Plateau, right/left, L, Uhmwpe recalled | Waldemar Link GmbH & Co. KG (Mfg Site) | Other | IN |
| Device | CLASS II | U2 Total Knee System Femoral offset bushing recalled | United Orthopedic | Device Malfunction | Nationwide |