CLASS II ONGOING Device recall · Reported 26 Aug 2026 · FDA Enforcement Report

Veritas Standard Reverse Total Shoulder Veritas Standard recalled

restor3d is recalling Veritas Standard Reverse Total Shoulder Veritas Standard Glenoid Baseplate, distributed nationwide in the US. The recall was initiated on 30 Jun 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: Model No 1500-1030; UDI-DI 00840097513225; Lot Numbers A00030, A00065, A09828, A09830 Model No 1500-1035; UDI-DI 00840097513232; Lot Numbers A00064, A09829
Quantity recalled59

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2987-2026
FDA event ID
99402
Recalling firm
restor3d, Durham, NC
Classification
Class II
Status
Ongoing
Recall initiated
30 Jun 2026
FDA classified
19 Aug 2026
Reported
26 Aug 2026
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

30 Jun 2026Recall initiated by company
19 Aug 2026Classified by FDA+50 days
26 Aug 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder

Where it was distributed

U.S. Nationwide distribution in the state of Texas.

Nationwide Texas

Questions about this recall

Why was it recalled?

Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.

Which lots are affected?

Lot Code: Model No 1500-1030; UDI-DI 00840097513225; Lot Numbers A00030, A00065, A09828, A09830 Model No 1500-1035; UDI-DI 00840097513232; Lot Numbers A00064, A09829

Where was it sold?

U.S. Nationwide distribution in the state of Texas.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from restor3d

All 5

Other Hip & Knee Implants recalls

All 3,111